Setting Up the Print Form Using PDM
Variables Available for the final Evaluation
80
FISCHERSCOPE
®
X-RAY
of the current product.
@SA1Standard deviation s_a.
This standard deviation expresses the additional distribution
of the mean values of fixed blocks that cannot be explained
by the standard deviation s of the single readings within the
blocks. Thus, it is an additional distribution that is based on
causes different from the distribution of the single readings.
This characteristic statistical quantity is available only with
fixed block sizes.
@SD1Standard deviation s of the single readings of the selected
blocks in the current product.
The standard deviation s is a measure for the distribution of
the single readings of the measurement series around their
common mean value. It equals the mean square deviation of
the single readings from the mean value and is computed
according to the equation found in the appendix “Statistical
formulas of measurement technology for the
FISCHERSCOPE X-RAY“.
For fixed block sizes @SD1 is the standard deviation of the
block mean values.
Regarding the numerals
1; 2; 3 … after certain
variables:
A measurement consists
of one or more single
readings (“channels”) that
are identified by consecu-
tive numbers (1,2,3).
Example “Dual coating
measurement”
Top coating thickness = 1
Intermediate coating
thickness = 2
@SDTDate of the first block of the selected blocks in the current
product. The date of the last measurement of the block
is used as the date for this block.
@SP1Statistical process control chart (SPC chart) for the
attribute 1 of the sample.
@STBNumber of the first block of the selected blocks in
the current product.
@STMTime of the first block of the selected blocks in
the current product. The time of the last measurement of
the block is used as the time for this block.
@UE1Lower control limit LCL(d) of the mean values of an
SPC chart.
@UG1Lower specification limit LSL (lower tolerance) for the
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......