Pattern Recognition (XDLM 237)
12.Define the search area:
Position of the search area: Place the cursor in the rectangle, keep the left mouse button
pressed and move the rectangle.
Size of the search area: Place the cursor on one of the eight black points, keep the left
mouse button pressed and change the size of the rectangle.
For the measurement of the solder pads, select the area of the vicinity of the pad as the search
area; see Fig. 21.5.
Figure 21.6: The solder pads on an individual printed circuit board
For the printed circuit board panel, a total of 10 pads are to be measured on each individual
printed circuit board; see Fig. 21.6. Of these, 6 pads are very similar. Only one search photo
is defined for them. The 4 lower pads are very different; one photo is defined for each of these.
13.In the
Move to saved xy-coordinates
area, click
Next position
.
The XY-stage moves to the next measuring position, to the next pad that is very similar to the
first one. For this measuring position, a separate search photo is not defined but the first search
photo is used.
14.To assign the search photo of the previous measuring position for this measuring position, enter
1 in the
Use picture no.
field and click
Change
.
Search image no. 1 is now used for XY position no. 2.
15.Repeat steps 13 to 14 for all 6 pads that are very similar.
16.In the
Move to saved xy-coordinates
area, click
Next position
.
The XY-stage moves to the next measuring position, to the next pad. This pad and the following
three pads are different. For this reason, a separate search image is defined for each of these
four pads.
17.Repeat steps 6 to 13 for each of the four different pads. During step 13, make sure that you
define a separate search photo for each pad this time.
18.You have now defined the 10 measuring positions for an individual printed circuit board on
the printed circuit board panel. To define the measuring positions for the other individual
printed circuit boards, you have two possibilities:
You can transfer the programmed search images to the other individual printed circuit
boards with the pattern recognition: Continue with step 19.
You can automate the measurement with a task: Continue with Chap. 21.4.1 on
page 182.
4 different pads
6 similar pads
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......