FISCHERSCOPE
®
X-RAY
155
Solution Analysis
19
Solution Analysis
For the solution analysis you need a measuring cell. You fill the measuring cell with the solution to
be analyzed, position the measuring cell and perform the measurement.
The measuring cells are available with different reference plates. When selecting the measuring
cell, note that the characteristic x-ray lines of the sample to be measured may not be too close to
the x-ray lines of the reference plate material. Otherwise, problems may arise during the evaluation
of the spectrum.
Table 19-1: Accessories for the solution analysis
19.1 Structure of the Measuring Cell
Figure 19.1:Structure of a measuring cell for the solution analysis
Order No.
Name
603-216
Accessory set for solution analysis Molybdenum: 2 measuring cells (red) with Mo-
lybdenum reference plate, 1 roll of foil
601-564
1 roll of foil
603-215
Measuring cell (red) with clamping ring, with Molybdenum reference plate
603-213
Measuring cell (green) with clamping ring, with Nickel reference plate
603-214
Measuring cell (grey) with clamping ring, with Zirconium reference plate
1 Bottom part of measuring cell
2 Edge of cup
3 Cup
4 Reference plate
5 Clamping ring
6 Drip groove
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......