Trueness of measurement readings
Measurement device monitoring for the Fischerscope X-RAY
FISCHERSCOPE
®
X-RAY
111
Variations
The variations of an X-ray fluorescence instrument depend on the
measurement parameters measurement time, distance between
the detector and the specimen, intensity of the primary beam and
other characteristics of the instrument as well as on the physics of
the X-ray fluorescence with regard to the measurement application.
It is interesting that with the X-ray fluorescence method, the
variations of the readings can be calculated theoretically from the
spectrum of a single reading such that a measure for the variation
can be specified with one single reading. The standard deviation is
used to evaluate the
repeatability precision
of the instrument with
regard to the measurement application, the feature to be measured
and the setting parameters of the instrument. The better the
repeatability precision of the instrument, the less the measurement
results will be ”falsified” by the instrument. Thus, a good
repeatability precision is considered an important quality feature of
an instrument and is of particular importance when evaluating the
suitability of a measurement device
.
Guaranteed Error
The calibration standards are subject to a ”measurement
uncertainty” as well because their nominal values are obtained
through measuring. For this reason, the nominal values of the
standards also include an error, the
guaranteed error
(e.g., 5% of
the nominal value) that includes the mean value, and its interval
should include the true value.
15.4 Trueness of measurement readings
The repeatability precision of an instrument must not be confused
with the
trueness
of a reading. The trueness of a measurement
result is determined by comparing the mean value B of a test series
of a sample, for example a calibration standard or a reference
sample, with its known value A that has been
recognized as being
correct
, the nominal value. The trueness is represented by the
difference B – A between the mean value and the nominal value.
The measurement result is correct if the difference B – A is near
Zero.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......