Pattern Recognition (XDLM 237)
Replay at Pos.No.
If the specimen has repetitive structures (e.g. a panel with sev-
eral individual printed circuit boards), then you can use this to
define the position at which the pattern recognition is to start
again from the beginning.
Max. rotation
If the measuring item is arranged so that it is skewed in rela-
tion to the defined search item (e.g. due to skewed posi-
tioning), then the measuring item can still be detected. Enter
the maximum expected rotation angle here; it should not be
larger than necessary though, because this increases the cal-
culation time.
Previous angle
Setting: If you have measuring items with sev-
eral measuring positions that could have a skewed position.
Set the angle for the first measuring point, e.g. +/- 120
degrees, and
Previous angle
for the other measuring points.
This saves calculation time.
Angle=
Shows the actual rotation angle that was found; see also
Max. Rotation
.
Pattern definition
Total video area
When activated, the search item is searched for in the entire
video image.
Show search item
Shows the search item (the image that is to recognized). Acti-
vate this option to define the search item.
Show search area
Shows the search area (the area of the video image in which
the search item is searched for). Activate this option to define
the search area.
Show cross hair
Shows the crosshairs of the video image.
Saved pictures
Deletes the current reference image with the corresponding
parameters or all reference images and all parameters.
Clear current
Clear all pict.
Correction / X / Y
When the found position is not exactly in the crosshairs, you
can correct it here. Entry in pixels. The correction has a global
effect on the instrument and not just the product.
Setting/parameter
Meaning
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......