User Interface of the WinFTM Software
To display the spectrum as colored area or colored line
►
In the window
Spectrum
click
. With this button you can toggle between both
representations.
To display the spectral lines of all elements of the current product
►
Select in the
Spectrum
window
View > Elements of Product
.
6.8.1 Saving a Spectrum
You can use a saved spectrum to re-evaluate it with a different measurement application, without
the need of performing the measurement again.
To save a a single spectrum
You can save a single spectrum in the WinFTM main window or in the
Spectrum
window.
1. Depending on the window that is currently opened, do one of the following:
•
The
Spectrum
window is opened: Select
Spectrum > Save as…
.
•
The WinFTM main window is opened: Select
General > Save Spectrum…
.
The window
Save as
appears. A default file path is specified, which contains the current high
voltage and the current primary filter, if available. Using this file path makes it easy to retrieve
the spectrum later.
2. Enter a file name and click
Save
.
3. The file will be saved with the extension
*.spk
or
*.spe
.
To save the spectrum after every measurement automatically
►
In the WinFTM main window select
Store Spectrum Automatically
. If the automatic storing is
activated, a disk symbol appears on the
Start
button.
Repeat the step to deactivate the automatic storing.
6.8.2 Loading a spectrum
You can load a saved spectrum for evaluation purposes at any time. You can also load several
spectra to compare them.
1. In the window
Spectrum
click .
The window
Open
appears.
2. Select the spectrum and click
Open
.
If a spectrum is already displayed in the spectrum window, then the newly loaded spectrum will be
displayed in a different color.
6.8.3 Deleting a spectrum
With this function you can remove one spectrum or several spectra from the spectrum window. It
does not make a difference whether you have just generated the spectrum by a measurement or
you have loaded a saved spectrum.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......