154
FISCHERSCOPE
®
X-RAY
Calibration, Base Correction and Reference Measurement
Procedure with manual positioning
1. Position the pure element silver (Ag) and close the measuring chamber.
2. Adjust the focus.
4. Click
Start Reference
.
The instrument performs the measurements. After it you will be prompted to position the pure
element copper (Cu).
5. Position the pure element copper (Cu) and close the measuring chamber.
6. Click
Start Reference
.
The instrument performs the measurement. When the measurement is finished, the window
Reference Measurement OK, Accept?
appears.
7. Click
OK
.
The measured values are stored.
3. In the WinFTM main window select
General > Reference Measurement
.
The window
Reference Measurement
appears.
The window displays date,
time, offset, gain and
control voltage of the
previous reference meas-
urement.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......