Calibration, Base Correction and Reference Measurement
d Click
Stop
.
e
Click
Exit
.
Procedure with automatic positioning
1. Position the pure element silver (Ag) and close the measuring chamber.
2. Adjust the focus.
4. Click
XY-Pos. „Ag“
.
The coordinates of the pure element silver are stored.
5. Position the pure element copper (Cu) and close the measuring chamber.
6. Click
XY-Pos. „Cu“
.
The coordinates of the pure element copper are stored.
7. Click
Start Reference
.
The instrument performs the measurements. When the measurements are finished, the window
Reference Measurement OK, Accept?
appears.
8. Click
OK
.
The measured values are stored.
3. In the WinFTM main window select
General > Reference Measurement
.
The window
Reference Measurement
appears.
The window displays date,
time, offset, gain and
control voltage of the
previous reference meas-
urement.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......