148
FISCHERSCOPE
®
X-RAY
Calibration, Base Correction and Reference Measurement
Procedure
1. In the WinFTM main window select
Calibrate > Calibration Standard Set…
.
An empty window will
appear, if the measuring
application has not yet been
calibrated.
If the measuring application
has already been cali-
brated, check the values in
the window. If the values
are plausible, click
OK
and
2. Enter a name for the calibra-
tion standard set in the field
Label
(e.g., Cr/Ni/Cu Ser.
No. 13917).
3. Click
Input/Overwrite
.
The window
Standard No. 1
appears.
4. Enter the name of the
standard (e. g., AAMGW).
5. Enter the values of the cali-
bration standard. Click the
field three times in order to
enter a value.
The fields have the following meaning:
Value
Nominal value of the calibration standard (in the example coat-
ing thickness d1 (Zn) in μm coating thickness d2 (Ni) in μm, Cu
concentration in %)
u +/-
If an uncertainty for the reference value is stated on the certificate
for the calibration standard, enter this value here. If you do not
enter a value, then the following values will be used: for coating
thicknesses, 1% of the coating thickness; for concentrations, 1%
absolute.
Type
With multi-coating systems, it may be the case that existing interim
coatings act as base material of a calibration standard or that
coatings are not present. This setting is made here.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......