FISCHERSCOPE
®
X-RAY
Components
To select the aperture
1. In the WinFTM main window select
Product > Modify…
.
The window
Modify product
appears.
2. Select the tab
Application
.
3. Click
Collim.
.
4. Select the collimator and click
OK
.
2.5.5 Shutter
The shutter is used to keep the primary x-rays from entering the measurement chamber and the en-
vironment.
The shutter is part of the safety system. The shutter closes immediately, when you open the cover
of the measurement chamber or when a malfunction occurs. X-rays can then no longer reach the
measurement chamber and the environment.
2.5.6 Specimen Lighting
The specimen lighting ensures good visibility of the specimen in the video window. The lighting
elements are white LEDs that are located in the area of the aperture.
Use the slide control in the video window to adjust the intensity of the illumination.
2.5.7 Primary filter
The primary filter consists of a thin foil made of nickel or aluminum, which is located between the
x-ray tube and the specimen to be measured.
By using a primary filter, the composition of the primary X-radiation can be influenced such that it
is optimally suited for the excitation of the respective specimen for x-ray fluorescence. Thus, unde-
sirable components of the fluorescence spectrum can be avoided from the outset.
Using a primary filter is useful for certain applications, however, it is not required in most cases.
The user requires special physical knowledge for a case-by-case evaluation of this question.
The primary filter can be controlled using the software, if the WinFTM extension SUPER is installed.
To change the primary filter:
1. In the WinFTM main window select
Product > Modify…
.
The window
Modify product
appears.
2. Select the tab
Application
.
3. Click
Def.MA (local)
.
The window
Definition of measuring conditions…
appears.
4. In the upper right field select the desired primary filter and click
OK
.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......