PN A51387A
B
xxi
Introduction
Before You Start: Warnings and Cautions
Cautions
In this document, a yellow header box alerts you to a caution message:
CAUTION
A caution message describes a potentially hazardous situation, which if not
avoided, may result in minor or moderate injury. It can also be used to alert
against unsafe practices.
Electrical Cautions
•
Power must be largely free from interference. The Multisizer 4 is CE, UL and CSA compliant with
regards to electrical interference but some circumstances may benefit from a constant voltage
transformer or regulator.
Chemical Cautions
•
If Beckman Coulter ISOTON® II is used, the Multisizer 4 must be drained and flushed through
with distilled water before carrying out a sodium hypochlorite solution (bleach) disinfection
procedure.
•
If the Multisizer 4 has been subjected to a bleach disinfection procedure, the instrument must
be drained and flushed through with distilled water before refilling with ISOTON® II.
•
Never stand containers of liquids on top of the Multisizer 4. Repair of instruments damaged or
affected by spilled liquids will not be covered by any warranty or guarantee.
NOTE
The Multisizer 4 performs best when it has reached a steady working temperature. This
typically takes 15 minutes from powering on the Analyzer.
Summary of Contents for Multisizer 4
Page 12: ...xii List of Figures ...
Page 14: ...xiv List of Tables ...
Page 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Page 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Page 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Page 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...