PN A51387A
B
xvii
Introduction
Multisizer 4 Specifications
Displaying Multisizer 4 Data
The Multisizer 4 can display data graphically as number, volume, mass, and surface area size
distribution plots, or as tables of values versus size. Both display types can show data as differential
or cumulative frequencies. Full data reduction facilities are included for statistical and trend
analysis, with export and import capabilities to text and spreadsheet programs via disk files or the
Windows Clipboard. The Multisizer 4 can print results using any printer compatible with the
Windows platform and drivers installed on the system.
A report facility allows results to be combined with sample information in a user-selectable way to
produce a systematic reporting configuration. Report formats can be individually named for ease of
archiving and retrieval. Number and date formats are supported in both conventional as well as in
ISO 9276-1 forms.
Multisizer 4 Components
The Multisizer 4 system comprises an integrated fluid handling and pulse processing unit (the
Analyzer Module) and a personal computer with a monitor with a minimum resolution of
1024 by 768. A keyboard, mouse and an ink-jet printer are also part of the system.
Multisizer 4 Specifications
Multisizer 4 dimensions and weight, environmental conditions, and operational parameters are
listed in this section.
Dimensions and Weight
Weight, dimensions and power (excluding computer):
•
Unpacked weight:
45 kg (99 lb)
•
Width:
45 cm (18 in)
•
Depth:
60 cm (24 in)
•
Height:
60 cm (24 in)
•
Input Voltage within Set Ranges:
100–120 VAC; 230–240 VAC ± 10%; single phase
•
Supply Frequency:
47 to 63 Hz inclusive
•
Power:
Less than 55 VA (watts)
•
Fuse types:
240 V, IEC (5
×
20 µm), Time delay (T), 3.15 A
NOTE
The use of other fuse types may result in damage to the instrument.
Summary of Contents for Multisizer 4
Page 12: ...xii List of Figures ...
Page 14: ...xiv List of Tables ...
Page 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Page 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Page 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Page 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...