PN A51387A
B
3-34
Installing and Calibrating an Aperture Tube
Running a Concentration Control Sample
2.
In the Control Sample Options window:
a.
Select Run a Control Sample.
b.
If you would like the Multisizer 4 software to prompt you to run a control sample at regular
intervals:
•
Select Hours or Days for the intervals between reminders.
•
In the Every field, enter the number of hours or days between control sample runs.
c.
Click
(
OK
)
.
Running a Concentration Control Sample
To run a concentration control sample:
1.
Fill a clean Accuvette® ST with 20 mL electrolyte solution and 0.2 mL (200 µL) calibrator.
2.
Place the Accuvette® ST on the Analyzer platform and raise the platform. For more information
on preparing the Analyzer, see
Preparing the Analyzer for Sample Runs
3.
Select
Calibration > Run Concentration Control
on the Main Menu bar. The Concentration
Control window opens.