PN A51387A
B
3-35
Installing and Calibrating an Aperture Tube
Running a Concentration Control Sample
3
4.
In the Concentration Control window, enter the Lot Number, the Concentration Assay Value,
and the Expiration Date listed on the assay sheet.
5.
Click
(
Start
)
to begin the concentration control run.
6.
The Multisizer 4 software flushes the system and begins the run.
7.
When the run is complete, a second Concentration Control window opens and displays results.
The window displays a message that the concentration is low or high if the Analyzer
measurement deviates from the concentration control by 10% or more. If this occurs, contact
Beckman Coulter Service. For information on how to do this, see
on
page
.
8.