PN A51387A
B
3-18
Installing and Calibrating an Aperture Tube
Verifying the Calibration
3.
In the Select Aperture Tube window, click
(
Edit List
)
. The Edit Aperture Tube List window
opens.
4.
In the Edit Aperture Tube List window:
a.
In the Diameter/Serial Number/Kd/Comment box, click the row you want to edit.
b.
Enter the new mean Kd in the Kd field.
c.
Click
(
Update
)
.
d.
Click
(
OK
)
to return to the Select Aperture Tube window.
NOTE
If you enter a Kd value that is too low or too high for the expected operation, the software
will display a warning message.
5.
In the Select Aperture Tube window:
a.
In the Diameter/Serial Number/Kd box, click to highlight an Aperture Tube.
b.
Click
(
OK
)
to return to the Wizard.
6.
Click
(
Close
)
to exit the Wizard.
Verifying the Calibration
When you verify aperture calibration, the Multisizer uses the Kd (or the mean Kd, if you have
obtained 10 Kd values and calculated the mean) to verify that the instrument is measuring
calibrator bead size accurately. Unlike the aperture calibration procedure (page
), which
returns a calibration constant (Kd) for the aperture orifice, the verification procedure returns the
measured calibrator bead size.
Because the Analyzer uses constant current, the instrument holds the calibration for an indefinite
time, provided that the working conditions are stable and you are using the same aperture tube and
electrolyte combination. If working conditions are not stable, you can select an option to prompt
Summary of Contents for Multisizer 4
Page 12: ...xii List of Figures ...
Page 14: ...xiv List of Tables ...
Page 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Page 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Page 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Page 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...