PN A51387A
B
5-19
Setting View and Print Preferences
Creating a Preferences File
5
•
In the Custom Trend Points pane, set trend percentiles, if desired. d[X] values allow you to
divide the particle distribution of each run by percentage of particle size above and below a
certain diameter. For example, d10 divides the particle distribution by the diameter at which
10% of the particles are smaller, and 90% larger.
•
In the Pulse Averaging pane, select pulse data to be included when viewing pulses as average
pulse height or average pulse width.
Preferences: Export
Use Step 4 of the Create Preferences Wizard (Export) or the Export tab of the Edit Preferences
window to select statistics you want to export to another file type, such as a Microsoft Excel
spreadsheet.
To proceed from Step 3 to Step 4, click
(
Next
)
.
To export statistics to another file type:
1.
Select the statistics you want to export.
2.
In the Data Format pane, select the type of list formatting you want. Exported list items can be
separated by commas, tabs, or blank spaces, or converted into an HTML table.
3.
If you select HTML, select table formatting for the exported file in the HTML Format pane.
4.
In Export Extension, enter the three-letter extension for the destination file type (for example,
XLS for Microsoft Excel). The most common file types are CSV and TXT.
5.
In the Numeric Format pane, select the numbers format you want.
Summary of Contents for Multisizer 4
Page 12: ...xii List of Figures ...
Page 14: ...xiv List of Tables ...
Page 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Page 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Page 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Page 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...