PN A51387A
B
9-7
Troubleshooting
Viewing Advanced Multisizer Settings
9
2.
Select
SERVICE > Calibrate Waste Tank
. The Status Panel displays CALIBRATING WASTE TANK.
CAUTION
Beckman Coulter does not recommend that you use the options available in
Service Mode without the assistance of a trained Service Representative. To
contact a representative, see
Requesting Service and Technical Support
page
.
Vacuum Settings
The Waste Tank and Metering Pump use vacuums to move fluid through the system. In both cases,
the default vacuum is set to six inches of mercury (6.00" Hg) for most aperture sizes. If you are using
time or count control mode and an aperture tube larger than 560 µm, you may need to change the
Waste Tank vacuum setting.
CAUTION
It is not possible to change vacuum or Metering Pump flow settings unless you are
running Service Mode. Beckman Coulter does not recommend that you use the
options available in Service Mode without the assistance of a trained Service
Representative. To contact a representative, see
.
To view vacuum settings:
1.
Select
Run > Troubleshooting > Sample Delivery Settings
on the Main Menu bar. The Sample
Delivery Settings window opens.