4-12
Selecting Analysis Settings: SOM and SOP
Using a Standard Operating Method (SOM)
To copy manual settings:
1.
Select
SOP > Edit the SOM
on the Main Menu bar.
2.
In the Edit the SOM window, select the Stirrer tab.
3.
On the Stirrer tab, select Other in the Sample Beaker pane and enter a non-standard beaker size.
4.
In the Stirrer Position pane, select Manual.
5.
In the Sample Compartment, manually position the stirrer.
6.
If desired, change stirrer speed and direction.
7.
In the Current Settings pane, click
(
Copy
)
. The current stirrer angle appears in the Manual
field.
8.
Click
(
Apply
)
or
(
Save SOM
)
to save your settings. The stirrer will move after you click
(
Apply
)
.