PN A51387A
B
9-11
Troubleshooting
Viewing Advanced Multisizer Settings
9
Draining the System
CAUTION
Draining the system is recommended only when cleaning, shipping, or
repackaging the Analyzer. As fluid evaporates within the system, remaining
electrolyte deposits salt crystals that can obstruct fluid flow and reduce
instrument performance.
To drain the system:
1.
Select
Run > Drain System
on the Main Menu bar. A Beckman Coulter pop-up window directs
you to remove or empty the Electrolyte Jar.
2.
Remove or empty the Electrolyte Jar, then click
(
OK
)
in the pop-up window. The system drains.
To view system drain time:
Select
Run > Troubleshooting > Fill, Flush and Drain Settings
on the Main Menu bar. The Fill, Flush
and Drain Settings window opens and displays system Fill Time, Flush Time, and Drain Time in
seconds.
CAUTION
It is not possible to change the system drain time setting unless you are running
Service Mode. Beckman Coulter does not recommend that you use the options
available in Service Mode without the assistance of a trained Service
Representative (see
Requesting Service and Technical Support
, page
To adjust system drain time:
1.
Select
Run > Troubleshooting > Enter Service Mode
on the Main Menu bar.
NOTE
Entering Service Mode will allow you to change default settings in the Troubleshooting menu.
If you do not enter Service Mode, settings in the Troubleshooting menu will appear as read-only
values.
2.
Select
Run > Troubleshooting > Fill, Flush and Drain Settings
on the Main Menu bar.
Summary of Contents for Multisizer 4
Page 12: ...xii List of Figures ...
Page 14: ...xiv List of Tables ...
Page 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Page 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Page 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Page 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...