Aperture current range:
30 µA to 6000 µA in 0.2-µA steps
•
Aperture current accuracy:
± 0.4% of setting
•
Polarity error:
Less than 0.5%
•
Time mode range:
0.01 to 1000 seconds, selectable in 10-ms increments
•
Metering system:
Mercury-free wide range metering pump
•
Metered sample volumes (volumetric mode only):
Continuously selectable from 50 µL to 2000 µL
•
Volumetric pump accuracy:
Better than 99.5%
Regulatory Compliance
This software is FDA 21 CFR part 11 compliant.
Before You Start: Warnings and Cautions
Warnings
In this document, a red header box alerts you to a warning message.
Aperture
(Nominal
Diameter,
µm)
Diameter Range (µm)
Volume Range µm
3
or fL
Total
Standard
Extended
Total
Standard
Extended
20*
0.4–16
0.4–12
12–16
0.034–2.14
×
10
3
0.034–905
905–2.14
×
10
3
30*
0.6–24
0.6–18
18–24
0.113–7.24
×
10
3
0.113–3.05
×
10
3
3.05
×
10
3
–7.24
×
10
3
50
1.0–40
1.0–30
30–40
0.524–33.5
×
10
3
0.524–14.1
×
10
3
14.1
×
10
3
–33.5
×
10
3
70
1.4–56
1.4–42
42–56
1.44–92.0
×
10
3
1.44–38.8
×
10
3
38.8
×
10
3
–92.0
×
10
3
100
2.0–80
2.0–60
60–80
4.19–268
×
10
3
4.19–113
×
10
3
113
×
10
3
–268
×
10
3
140
2.8–112
2.8–84
84–112
11.5–736
×
10
3
11.5–310
×
10
3
310
×
10
3
–736
×
10
3
200
4.0–160
4.0–120
120–160
33.5–2.14
×
10
6
33.5–905
×
10
3
905
×
10
3
–2.14
×
10
6
280
5.6–224
5.6–168
168–224
92.0–5.88
×
10
6
92.0–2.48
×
10
6
2.48
×
10
3
–5.88
×
10
6
400
8.0–320
8.0–240
240–320
268–17.2
×
10
6
268–7.24
×
10
6
7.24
×
10
6
–17.2
×
10
6
560
11.2–448
11.2–336
336–448
736–47.1
×
10
6
736–19.9
×
10
6
19.9
×
10
6
–47.1
×
10
6
1000**
20–800
20–600
600–800
4189–268
×
10
6
4189–113
×
10
6
113.1
×
10
6
–268
×
10
6
2000**
200–1600
200–1200
1200–1600
4.19
×
10
6
–2.14
×
10
9
4.19
×
10
6
–905
×
10
6
905
×
10
6
–2.14
×
10
9
* Range depends on system cleanliness and environmental electrical noise.
** Range depends on sample density.
Summary of Contents for Multisizer 4
Page 12: ...xii List of Figures ...
Page 14: ...xiv List of Tables ...
Page 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Page 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Page 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Page 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...