Analyzer Overview
Starting the Analyzer
Starting the Analyzer
To start the Analyzer:
1.
Power ON the Analyzer.
2.
Empty the Waste Jar, if necessary.
3.
Fill the Electrolyte Jar with electrolyte, if necessary.
4.
Open the Multisizer 4 software and connect the software to the Analyzer.
5.
On the Multisizer 4 software Main Menu bar, select
Run > Instrument Start-Up
.
6.
In the Instrument Start-Up dialog box:
a.
Select checkboxes for each function you would like the Analyzer to perform on startup.
b.
Click
(
Start
)
.
NOTE
The Multisizer 4 performs best when it has reached a steady working temperature. This
typically takes 15 minutes from powering on the Analyzer.
Preparing the Analyzer for Sample Runs
To prepare the analyzer for a calibration, control, or sample analysis run:
1.
Start the Analyzer and connect the software to the Analyzer (see
,
page
2.
Wash the ST Aperture Tube and ST Beaker or Accuvette® ST (see
, page
).
3.
Fill an ST Beaker or Accuvette® ST with clean electrolyte solution.