Analyzer Overview
Sample Compartment
Figure 1.6
Platform in Loading Position: Accuvette® ST Side
Reversing the Sample Platform
To turn the sample platform from the beaker to Accuvette® ST side, or vice versa:
1.
Lower the platform to the loading position, if necessary.
NOTE
Rotate the platform only when it is located at or near the loading position. This ensures that the
platform does not bump the aperture tube or stirrer while rotating.
2.
Grasp the platform and lift the platform slightly. It is not necessary to use the platform release.
3.
Pull the platform toward you. You will feel the resistance of a spring mechanism.
4.
Continuing to hold the platform away from the rear of the sample compartment, turn the
platform until it contacts the stop pin (
).
5.
Slowly release the platform.