PN A51387A
B
3-11
Installing and Calibrating an Aperture Tube
Calibrating the Aperture
3
CAUTION
Sizing and statistical errors will occur if you do not use the proper procedure and/
or calibrator size.
Nano-Size Particle Calibration Standards
If you use polystyrene nano-particle calibrators, Beckman Coulter recommends that you keep the
concentration level below 2%.
Polystyrene nano-particles can bind to system components. After the Multisizer 4 has analyzed a
high concentration of nano-particles, it is necessary to clean the aperture tube and external
electrode thoroughly.
Systems with high levels of nano-particle contamination will require several filling and draining
cycles to return background counts to normal levels. For information on the recommended cleaning
procedure, see
Selecting Calibration Options
To select calibration options:
1.
Select
Calibration > Aperture Calibration Options
on the Main Menu bar. The Aperture
Calibration Options window opens.
NOTE
The
Calibration > Aperture Calibration Options
menu item is only available to user types with
sufficient security privileges. If the Multisizer 4 software is running with no security, every
operator can calibrate the aperture.
200
20 – 40
280
28 – 56
400
40 – 80
560
56 – 112
1000
100 – 200
2000
200 – 400
Table 3.2
Choosing Calibrator Size
Aperture Size (µm)
Calibrator Mean or
Modal Size (µm)
Summary of Contents for Multisizer 4
Page 12: ...xii List of Figures ...
Page 14: ...xiv List of Tables ...
Page 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Page 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Page 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Page 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...