2
Conductivity Methods
Systems using the conductivity method, such as the i-STAT System, measure the electrical conductance of
a whole blood sample. Plasma conducts electrical current, and blood cells act as insulators. A sample with
a relatively high hematocrit has, by definition, a larger proportion of its volume filled by the non-conductive
red blood cells. The overall conductance of the sample will thus be relatively low. In the i-STAT System,
before the measured sample conductance is converted into a hematocrit value, corrections are applied for
the temperature of the sample, the size of the fluid segment being measured, and the relative conductivity
of the plasma component. The first two corrections are determined from the measured value of the calibrant
conductance; the last correction is determined from the measured concentrations of sodium and potassium
in the sample.
i-STAT provides two customization settings for reporting hematocrit results: The “K
3
EDTA” customization
reports hematocrit results traceable to MH-K
3
EDTA. The “K
2
EDTA” customization reports hematocrit results
traceable to MH-K
2
EDTA.
For best agreement of i-STAT and hematology analyzer hematocrit results, the i-STAT customization
setting is selected according to the calibration of the comparative hematology analyzer (MH-K
2
EDTA or
MH-K
3
EDTA).
When this Technical Bulletin was first released, the manufacturers of the Advia
®
, Cell-Dyn
®
, Coulter
®
, and
Sysmex
®
, hematology analyzers used samples collected in K
3
EDTA for the purpose of calibrating their
systems to reference methods. The manufacturers of Advia, Cell-Dyn and Sysmex now use samples
collected in K
2
EDTA for the purpose of calibrating their systems to reference methods. Therefore, for best
agreement between these analyzers and the i-STAT System, the K
2
EDTA customization setting should be
selected. (Note: The default setting on the i-STAT System is K
3
EDTA.)
Automated Hematology Analyzer Methods
Automated hematology analyzers do not directly measure hematocrit, but rather calculate hematocrit from
measurements of individual cell sizes and counts. The sample is diluted with an isotonic buffered aqueous
solution and passed through a measuring orifice. Individual cells are counted and sized by one of two basic
mechanisms. Either fluctuation in electrical conductivity or the scatter of collimated light, is measured as
cells pass through the orifice. The hematocrit of the original sample is calculated from the number of cells,
the sizes of the cells and the volume of diluted solution passing through the orifice.
Co-oximetry Methods
Like hematology analyzers, co-oximeters do not directly measure hematocrit. Hematocrit values are
calculated from hemoglobin values that are measured by spectrophotometry. A fixed relationship between
hemoglobin and hematocrit is used to calculate hematocrit.
METHOD COMPARISONS
The discrepancy between the hematocrit values determined by two different systems on an individual
sample has three components: a random component, a systematic component and a method-dependent /
sample-specific component.
Random Component
The random component results from the combination of each system’s imprecision. The size of the
component is different on each determination. The imprecision of the i-STAT System is typically about
0.6 %PCV to 0.7 %PCV (1 standard deviation).
Systematic Component
The systematic component results from differences in the systems’ calibration, and is constant from sample
to sample.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......