MEDICAL HEMATOCRIT CALIBRATION VERIFICATION CONTROLS
Intended Use
RNA Medical Hematocrit Calibration Verification Control is a quality control
material used for verifying the calibration and linearity of hematocrit sensors in
i-STAT cartridges.
Product Description
Hematocrit Calibration Verification Controls are provided in five (5) levels for
verifying the accuracy of hematocrit measurements throughout the reportable
range. The controls are packaged in sealed glass ampules, each containing 1.7 mL
of solution. Ampules are packaged in kits containing four (4) ampules of each level.
Active Ingredients
This product is a buffered aqueous solution containing electrolytes and non-
conductive ingredients. This control contains no red blood cells, and no human
or biological material.
Storage
RNA Medical Hematocrit Calibration Verification Control is stable until the
expiration date stated on the ampule when stored at temperatures of 2-25 ºC. Do
not freeze or expose ampules to temperatures greater than 30 ºC.
If stored refrigerated, the calibration verification material should be equilibrated
to room temperature for at least four (4) hours prior to testing.
The product may turn slightly yellow in color over its shelf life if stored at
room temperature. This coloration is normal and does not affect product
performance.
Directions For Use
1. Beginning with Level 1, gently invert the ampule to mix the solution.
Tap the ampule to restore the liquid to the bottom of the ampule.
2. Open the ampule by snapping off the tip at the neck. Use gauze, tissue,
gloves, or an appropriate ampule opener (a Snapper is included for this
purpose) to protect fingers from cuts.
3. Fill and seal a cartridge and insert immediately into the analyzer.
i. Note: The CalVer option from the Quality Tests menu must
be used on the i-STAT 1 Analyzer for RNA Medical Hematocrit
Calibration Verification Control.
4. Record the results on the Data Collection and Linearity Worksheet
provided.
5. Test Levels 2, 3, 4, and 5 in the same manner. Record all values on the
worksheet.
i. Note: this product is a buffered aqueous solution containing
electrolytes and non-conductive ingredients. The only value
assigned to this fluid is for hematocrit. All other cartridge analyte
results obtained with this control material should be ignored.
Always ensure that the lot number and software revision on the value
assignment sheet matches the lot number of the vial in use and the software
revision in the analyzer.
Target values are specific to the i-STAT System. Results may differ if used with
other methods.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......