Rev. Date: 01-Aug-11
Art: 714179-00L
iCa - 3
Method Comparison (mmol/L)
Radiometer ICA1
Nova STAT Profile
n
47
57
Sxx
0.009
0.017
Syy
0.017
0.017
Slope
0.925
0.960
Int’t
0.113
0.062
Sy.x
0.035
0.029
Xmin
0.46
0.53
Xmax
2.05
2.05
r
0.982
0.982
Cartridge Comparison
The performance characteristics of the sensors are equivalent in all cartridge configurations. System
difference analysis was performed on 24 patient samples using i-STAT CHEM8+ and i-STAT CG8+
cartridges. In the 0.46 - 1.23 mmol/L range, the average difference was 0.003.
Factors Affecting Results*
Venous stasis (prolonged tourniquet application) and forearm exercise may increase ionized calcium due to
a decrease in pH caused by localized production of lactic acid
6
. Exposing the sample to air will cause an
increase in pH due to the loss of CO
2
which will decrease ionized calcium.
Heparin binds calcium. Each unit of heparin added per mL of blood will decrease ionized calcium by 0.01
mmol/L.
6
Therefore, the correct ratio of heparin anticoagulant to blood must be achieved during sample
collection. Intravenous injection of 10,000 units of heparin has been shown in adults to cause a significant
decrease of ionized calcium of about 0.03 mmol/L.
6
Use only unheparinized sample transfer devices when
using i-STAT’s aqueous control and calibration verification materials.
Hemodilution of the plasma by more than 20% associated with priming cardiopulmonary bypass pumps,
plasma volume expansion or other fluid administration therapies using certain solutions may cause clinically
significant error on sodium, chloride, ionized calcium and pH results. These errors are associated with
solutions that do not match the ionic characteristics of plasma. To avoid these errors when hemodiluting by
more than 20%, use physiologically balanced multi-electrolyte solutions containing low-mobility anions (e.g.
gluconate) such as Normosol
®
-R (Abbott Laboratories), Plasma-Lyte
®
-A (Baxter Healthcare Corporation),
and Isolyte
®
-S (B Braun Medical) rather than solutions such as normal saline or Ringer’s Lactate.
Test concentrations used were as per the CLSI guidance document,
7
unless otherwise indicated.
The following substances are known to interfere with the i-STAT Ionized Calcium assay:
Substance
Test Concentration
(mmol/L)
Interference
Acetaminophen
1.32
Decreased i-STAT Ionized Calcium results.
See Note below.
Acetylcysteine
10.2
Decreased i-STAT Ionized Calcium results.
See Note below.
Bromide
37.5
Increased i-STAT Ionized Calcium results.
See Note below.
Magnesium
1.0
Increased i-STAT Ionized Calcium results by
approximately 0.04 mmol/L.
Lactate
6.6
Decreased i-STAT Ionized Calcium results by
approximately 0.07 mmol/L.
Salicylate
4.34
Decreased i-STAT Ionized Calcium results.
See Note below.
Salicylate
(therapeutic)
0.5
10
Decreased i-STAT Ionized Calcium results by
approximately 0.03 mmol/L. See Note below.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......