When the calibration of a comparative method is uncertain, determine the customization setting by
minimizing the average bias between methods as follows:
• Check that the results from hematocrit controls for both i-STAT and comparative methods are
acceptable.
• If i-STAT hematocrit results obtained using the “
K
3
EDTA
” setting are consistently lower than those
on the comparative method, the “
K
2
EDTA
” setting may be a better choice. If agreement is better
after multiplying the “
K
3
EDTA
”-customized i-STAT results by 1.0425, the customization setting
should be switched to “
K
2
EDTA”.
• Conversely, if i-STAT hematocrit results obtained using the “
K
2
EDTA
” setting are consistently
higher than those on the comparative analyzer, the “
K
3
EDTA
” setting may be a better choice.
If agreement is better after dividing the “
K
2
EDTA
”-customized i-STAT results by 1.0425, the
customization setting should be switched to “
K
3
EDTA”.
• If an unacceptable system bias still exists, contact i-STAT Technical Support at 1-800-366-8020,
option 1.
Hematology analyzers anD K
2
eDta anD K
3
eDta samPle collection tubes
Hematocrit results on hematology analyzers from samples collected in K
3
EDTA and K
2
EDTA tubes will be
equivalent. This is because the osmotically-balanced diluent reverses the red blood cell shrinkage caused
by the anticoagulant.
5
It should be clear that results from K
2
EDTA and K
3
EDTA tubes will be equivalent, but
lower, on an analyzer calibrated to MH-K
3
EDTA than on an analyzer calibrated to MH-K
2
EDTA.
i-STAT has become aware that some customers have selected their i-STAT hematocrit customization
according to the type of EDTA anticoagulant in the collection tube used for samples for the hematology
analyzer. As explained above, the selection of the “K
2
EDTA” or the “K
3
EDTA” customization for i-STAT
analyzers is based upon the microhematocrit method (MH-K
2
EDTA or MH-K
3
EDTA) to which the hematology
analyzer is calibrated, rather than on the collection tube used for the hematology analyzer.
exPecteD level of metHoD agreement
Average i-STAT hematocrit results over a group of samples should normally agree with those from the
comparative method within ± 2 %PCV at 29 %PCV and below, ± 3 %PCV from 30 to 50 %PCV, and within
10% above 50 %PCV when the following conditions are met:
• i-STAT analyzers are customized correctly.
• Comparative analyzer is calibrated correctly.
• Sample handling is optimal for both i-STAT and comparative methods.
• Samples are unaffected by factors listed in the i-STAT Cartridge and Test Information sheet for
Hematocrit or in the user documentation for the comparative method.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......