![background image](http://html1.mh-extra.com/html/abbott/i-stat-1/i-stat-1_system-manual_450958405.webp)
CARTRIDGE AND TEST
INFORMATION
Art:
714258-00N
Rev. Date: 22-May-12
i-STAT sensors are available in a variety of panel configurations. Sensors are contained in cartridges with microfluidic
components and, in some cartridges, calibration solution. i-STAT cartridges are used with the i-STAT Portable Clinical
Analyzer and the i-STAT 1 Analyzer* for the simultaneous quantitative determination of specific analytes and coagulation
parameters in whole blood.
CARTRIDGE SPECIFICATIONS
Shelf Life:
Refrigerated at 2 to 8ºC (35 to 46ºF) until expiration date.
Refer to the cartridge box for room temperature storage requirements.
Preparation for Use:
Individual cartridges may be used after standing five minutes at room temperature. An
entire box of cartridges should stand at room temperature for one hour.
All cartridges should be used immediately after opening pouch. If the pouch has been
punctured, the cartridge should not be used.
Sample Type:
Fresh whole blood from arterial, venous, or skin punctures
(Note: Skin puncture is NOT a recommended sample type for ACT, cTnI, CK-MB, or BNP
testing.)
cTnI and CK-MB cartridges require the use of heparinized whole blood or plasma, or non-
heparinized whole blood tested within one minute of patient draw.
BNP cartridges require the use of EDTA whole blood or plasma samples.
Sample Volume:
17
µ
L, 20
µ
L, 40
µ
L, 65
µ
L, or 95
µ
L depending on cartridge type.
Test Timing:
Immediately after collection
•
Samples for the measurement of ACT, PT/INR and Lactate
Within 3 minutes after collection
•
Samples collected in capillary tubes, both with and without anticoagulant
•
Samples collected in evacuated or non-evacuated tubes and syringes without
anticoagulant
Within 10 minutes after collection
•
Samples collected with anticoagulant for the measurement of pH,
P
CO
2
,
P
O
2
,
TCO
2
and iCa. Maintain anaerobic conditions. Remix before filling cartridge.
Within 30 minutes after collection
•
Samples collected with anticoagulant for the measure of Sodium, potassium,
chloride, glucose, BUN/urea, creatinine, hematocrit, troponin I, CK-MB, and
BNP. Remix thoroughly before testing.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......