CALIBRATION VERIFICATION
15
Art: 714377-00K
Rev. Date: 31-Jan-12
15-1
CALIBRATION VERIFICATION FOR BLOOD GAS/ELECTROLYTE/METABOLITE CARTRIDGES
(EXCEPT CHEM8+ CARTRIDGES)
Purpose
Calibration Verification is a procedure intended to verify the accuracy of
results over the entire measurement range of a test. The performance of this
procedure at defined intervals may be required by regulatory or accreditation
bodies. While the Calibration Verification Set contains five levels, verification
of the measurement range could be accomplished using the lowest, highest
and mid levels.
Overview of Procedure
i-STAT recommends that each sensor type be included in the Calibration
Verification procedure using a selection of analyzers that have passed the
Electronic Simulator check. See the Technical Bulletin "Calibration Verfication
and the i-STAT System" for more information.
Calibration Verification
Solutions for
Cartridges
A five-level Calibration Verification Set is available to verify the calibration
of i-STAT cartridges throughout the reportable ranges for:
Sodium
pH
Glucose
Potassium
P
CO
2
Lactate
Chloride
P
O
2
BUN/Urea
Ionized Calcium
Creatinine
There are four 1.7mL glass ampules of each level in the set.
Storage
Refrigerated storage at 2 to 8 °C (35 to 46°F) should be maintained until the
printed expiration date on the box and ampule labels. Calibration Verification
fluids may also be stored at room temperature for up to 5 days (18 to 30 °C or
64 to 86°F). Prolonged storage at temperatures greater than 30 °C (86°F) may
cause changes in the values of some analytes. Do not use beyond the expiration
date on the box and ampule labels.
If stored refrigerated, the calibration verification material should be equilibrated
to room temperature for at least four (4) hours prior to testing.
Ampule Use
When using cartridges that contain sensors for pH,
P
CO
2
,
P
O
2
and ionized
calcium, a separate ampule must be used for each cartridge being tested. If
these sensors are not present, the contents of one ampule may be used to fill
more than one cartridge as long as the cartridges are filled and inserted into
an analyzer within 10 minutes of opening the ampule.
Best Results
For best results, ampules, cartridges and analyzers should be at the same
temperature.
Reactive Ingredients
See the table on page 14-3 of the Quality Control section for full information.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......