Rev. Date: 01-Mar-12
Art: 715236-00N
PT/INR - 3
The below imprecision data for lyophilized plasma control material were collected during studies at an
Abbott Point of Care facility and during clinical trials. SD and %CV are typical of current performance.
Current Value Assignment Sheets should be referenced for applicable plasma control mean data.
Plasma Control
Mean
SD
%CV
Level 1
1.1 (INR)
0.05
4.5%
Level 2
2.5 (INR)
0.17
6.9%
Reference Interval
In a study to determine a reference interval for PT/INR, venous samples from healthy volunteers were
collected in plastic tubes, and whole blood was analyzed with one lot of cartridges on the i-STAT System.
Capillary samples were obtained from the same volunteers using Softclick Pro (setting of 3) and analyzed
on the same cartridge lot. Reference intervals for INR in venous and capillary samples were determined
according to the CLSI Guideline C28-A2.
3
The data are summarized in the table below:
Statistic
Venous whole blood
Capillary whole blood
n
120
119
Mean (INR)
1.0
1.0
SD
0.1
0.1
Reference Range (INR)
0.8 - 1.2
0.8 - 1.2
Due to the many variables that may affect PT/INR results, each laboratory should establish its own
reference interval.
Method Comparison
Method comparison data were collected at the Hemostasis Reference Laboratory (Hamilton, Ontario,
Canada). Venous samples from outpatients undergoing routine oral anticoagulation therapy were collected
in plastic tubes and analyzed in duplicate on multiple lots of cartridges on the i-STAT System; plasma from
tubes containing a citrate anticoagulant were analyzed in duplicate on the comparative instruments using
Dade
®
Innovin
®
, STA Neoplastine
®
CI Plus, and the HemosIL
®
RecombiPlasTin 2G
®
reagents.
Deming regression analysis
4
was performed on the first replicate of each sample. In the method
comparison table below,
n
is the number of specimens in the data set,
Sy.x
is the standard error of the
estimate, and
r
is the correlation coefficient.
Method comparisons will vary from site to site due to differences in the sample handling, reagent and
instrument systems in use, and other site-specific variables. A correlation study should be performed to
establish the differences between the i-STAT PT/INR measurement and other methods used.
Statistic
i-STAT vs. Siemens Sysmex
®
CA-1500 and Dade
®
Innovin
®
reagent
i-STAT vs. STA Compact
®
and
Neoplastine
®
CI Plus reagent
i-STAT vs. IL ACL 7000 and
HemosIL
®
RecombiPlasTin
2G
®
reagent
n
78
78
69
Mean (INR)
2.1
2.1
2.2
Range (INR)
0.9 - 4.5
0.9 - 4.1
0.9 – 4.0
Sx (INR)
0.843
0.772
0.840
Slope
0.981
1.074
0.972
Intercept (INR)
0.084
-0.100
0.003
r
0.963
0.964
0.962
Sy.x
0.233
0.229
0.233
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......