cTnI - 6
Art:
715595-00R
Rev. Date: 23-May-12
Test Limitations
The frequency of suppressed results is affected by atmospheric pressure. Suppressed result rates may
increase with higher elevations (decreased barometric pressure) and may become persistent if testing is
performed at more than 7500 feet above sea level. Where unavailability of results is unacceptable, i-STAT
recommends having an alternate test method available.
Samples from patients who have been exposed to animals or who have received therapeutic or diagnostic
procedures employing immunoglobulins or reagents derived from immunoglobulins may contain antibodies,
e.g. HAMA or other heterophile antibodies, which may interfere with immunoassays and produce erroneous
results.
18-24
The generation of potentially interfering antibodies in response to bacterial infections has been
reported.
16
While this product contains reagents that minimize the effect of these interferents, and QC
algorithms designed to detect their effects, the possibility of interference causing erroneous results should
be evaluated carefully in cases where there are inconsistencies in the clinical information. Results from the
i-STAT cTnI assay should be considered in the context of the entirety of the available clinical information.
Medical decisions should not be based on a single i-STAT measurement.
14
Cardiac troponin may not appear in circulation for 4-6 hours following the onset of symptoms of MI.
Consequently, a single negative result is insufficient to rule out MI. The use of a serial sampling protocol is
recommended practice.
11
The results of different troponin assays are not generally comparable: cTnI and cTnT are distinct molecules
and results are not interchangeable, nor comparable. In addition, significant variation in absolute troponin
values may be observed for a given patient specimen with different analytic methods.
13
Partially clotted samples can result in elevated cTnI results above the reference range, as well as quality
check code errors. To prevent this from occurring, upon drawing the whole blood sample into a heparinized
collection tube, the sample should be inverted gently at least 10 times to ensure even dissolution of the
heparin anticoagulant.
Grossly hemolyzed samples can cause a decreased alkaline phosphatase activity, resulting in decreased
detection of cTnI, increased assay backgrounds, and/or quality check codes.
Hematocrits in the range of 0-65 % PCV have been demonstrated not to affect results. Samples with
hematocrit levels above this range have demonstrated increases in the test imprecision and quality check
codes.
The analyzer must remain on a level surface with the display facing up during testing. Motion of the analyzer
during testing can increase the frequency of suppressed results or quality check codes. A level surface
includes running the handheld in the downloader/recharger.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......