StabIlIty of calIbratIon In the i-Stat SyStem
The i-STAT System is a unit-use testing system. Components that cause shifts and drifts in results in multi-
use analyzers: sensors (electrodes), calibration solution, fluid-handling channels and pumps, are housed in
a disposable test cartridge. The sensors are exposed to sample only once, so there is no protein build-up
which is a major cause for deterioration of sensor slope and the need to calibrate and/or verify calibration
on a frequent basis in multi-use analyzers.
The stability and consistency of the manufacturing process allow the slope of the sensors to be
programmed into the analyzer’s software. A one-point calibration to set the intercept accounts for any day-
to-day variation in testing conditions. When stored according to directions, the cartridges are stable up to
the expiration date.
The analyzer houses the mechanical and electrical systems necessary to control fluid movement within
the cartridge, control the temperature when measurements are performed at 37°C, measure barometric
pressure, measure electrical signals generated by the sensors and display and transmit results. The
analyzer’s functions are factory calibrated to specifications that are programmed into the analyzer along
with acceptability limits, which when exceeded cause the analyzer to display quality check messages or to
display *** rather than results.
The accuracy of results and dependability of the internal quality check system depend upon the ability
of the analyzer to take accurate and sensitive signal readings from the sensors. To check this function,
i-STAT developed an electronic control device. The Electronic Simulator simulates two levels of electronic
signals that stress the analyzer’s signal detection function both below and above the reportable ranges.
Injecting signals directly into the analyzer allows very tight control limits to be set. Control limits for liquid
controls are set wide enough to allow for sensor-to-sensor variation. All analyzers that pass the Electronic
Simulator test are equivalent and any variations in results are caused by within and between lot variations in
the cartridges.
The combination of unit-use cartridges, inherently stable electronics of the analyzer, and reliability of the
Electronic Simulator check provides the stability needed for a point-of-care testing system and reduces the
need for frequent stability or calibration verification checks.
ValIdatIng the reportable range
The accuracy of results over the entire reportable range could be assessed by testing the same patient
samples on the new system and on a system with known accuracy and comparing results using an
acceptable difference criteria. However, it is difficult to find samples that cover the low and high ends of the
reportable ranges of many analytes. Ideally, the samples used to validate the ranges should have the same
matrix as the patient samples. Tonometered blood can be used to validate the ranges for
P
CO
2
and
P
O
2
.
But for most analytes, there are no whole blood samples commercially available for this purpose.
Although the i-STAT Calibration Verification Set is aqueous-based, the target values have been determined
over many lots of cartridges and results on these solutions when compared to the target values should
indicate the performance of a particular lot of cartridges.
aSSeSSIng lot-to-lot VarIatIon
Lot-to-lot variation over the entire reportable range for any reagent system could be assessed by testing
calibration verification solutions on old and new lots in parallel. This procedure should be suspended once
it has been determined that lot-to-lot variation is acceptable. Quality controls samples with concentrations
at decision points should always be used to assess new lots or reagent before results are reported.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......