PROCEDURE MANUAL FOR THE i-STAT SYSTEM
1
REV. DATE: 16-Oct-12
ART: 714446-00L
SYSTEM OVERVIEW
The i-STAT System incorporates comprehensive components needed to perform blood analysis at the point of care. The
System consists of the following primary components:
Analyzers
Analyzers (handhelds) can be the i-STAT Portable Clinical Analyzer or the i-STAT 1 Analyzer. When a sample-filled
i-STAT cartridge is inserted into a handheld for analysis, the handheld automatically controls all functions of the testing
cycle including fluid movement within the cartridge, calibration and continuous quality monitoring.
Analysis Time
ACT cartridge: to detection of end point - up to 1000 seconds (16.7 min.)
PT/INR cartridge: to detection of end point – up to 300 seconds (5 min.)
cTnI and BNP cartridges: 600 seconds (10 min.)
CK-MB Cartridge: 300 seconds (5 min.)
Other cartridges: typically 130 to 200 seconds
Cartridges
A single-use disposable cartridge contains microfabricated sensors, a calibrant solution, fluidics system, and a waste
chamber. Sensors for analysis of pH,
P
CO
2
,
P
O
2
, TCO
2
, sodium, potassium, chloride, ionized calcium, glucose, lactate,
creatinine, urea nitrogen (BUN) and hematocrit are available in a variety of panel configurations. Cartridges are also
available for Celite-ACT, Kaolin-ACT, PT/INR, Troponin I/cTnI, CK-MB and BNP (Table 1). A whole blood sample of
approximately 1 to 3 drops is dispensed into the cartridge sample well, and the sample well is sealed before inserting it into
the analyzer.
Central Data Station or Data Manager
A dedicated desktop computer with the i-STAT Central Data application provides the primary information management
capabilities for the i-STAT System. IR Links for Portable Clinical Analyzers and Downloaders and Downloader/Rechargers
for the i-STAT 1 Analyzers allow for transmission of patient records from a widely distributed network of handhelds to the
Central Data Station application. Data can be stored, organized, edited, and transferred to a laboratory information system or
other computer system. Cartridge usage and efficiency reports can be generated for management of the System.
SUPPLIES and STORAGE REQUIREMENTS
Cartridges
Cartridges are sealed in individual pouches or portion packs. Store the main supply of cartridges at a temperature between 2
to 8°C (35 to 46°F).
Do not allow cartridges to freeze.
Cartridges may be stored at room temperature (18 to 30°C or 64 to
86°F) for the time frame indicated on the cartridge box. Cartridges should not be returned to the refrigerator once they have
been at room temperature, and should not be exposed to temperatures above 30°C (86°F). If the pouch has been punctured,
the cartridge should not be used. Write the date on the cartridge box or individual cartridge pouches to indicate the room
temperature expiration date. Cartridges should remain in pouches until time of use. Do not use after the labeled expiration
date.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......