Art: 714364-00M
Rev. Date: 02-Aug-12
2-13
Customization
Analyzers can be customized for site-specific testing characteristics and
requirements. A complete list of customizable parameters and their default values
can be found in the Customization section. An analyzer can be customized via the
keypad or via the Central Data Station. Items that cannot be customized via the
analyzer’s keypad are operator lists, test strip lists, reference and action ranges,
sample types and order of items on the Chart page.
The Central Data Station’s Customization function can be used to create one
customization profile for all analyzers or different profiles for different locations.
When the Customization function is enabled, the profiles are transmitted to the
analyzers when they are placed in a Downloader.
Caution: If location specific customization profiles are created, analyzers should
not be moved from one location to another unless they are re-customized for
the new location. This is especially important if “CPB: Automatically Adjust” or
“CPB: Do Not Adjust” is included in a location-based customization profile. The
CPB function adjusts hematocrit and hemoglobin results for the dilutional affect of
pump fluid during cardiopulmonary bypass surgery. If an analyzer customized for
the CVOR as “CPB: Automatically Adjust” is used for patients who are not on the
pump, hematocrit results will be reported falsely high. If an analyzer customized
as “CPB: Do Not Adjust” is used for patients who are on the pump, hematocrit
results will be reported falsely low. For details on the CPB function, see the Theory
section of this manual.
It is recommended that only one method, the Central Data Station or the keypad,
be used to customize all analyzers within a site. If both methods are in use, and the
Customization function is not disabled on the Central Data Station, any changes
made to the profile of an analyzer via the keypad will be overwritten the next time
the analyzer is placed in a Downloader.
The customization profile of an analyzer is identified in the Customization option
under the Administration Menu on the analyzer. DEFAULT0 indicates that the
analyzer has factory settings. When an analyzer has been customized via the
Central Data Station (CDS), the name assigned to the profile by the CDS is listed.
If the default or CDS profile is changed on the analyzer, the profile is listed as
00000000.
Note: The i-STAT Portable Clinical Analyzer can be customized only from the Central
Data Station. However, not all customizable features apply to this handheld. See
the i-STAT System Manual for the i-STAT Portable Clinical Analyzer for customizable
features for this handheld.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......