CELITE ACTIVATED CLOTTING TIME/
(
CELITE
ACT)
Rev. Date: 30-Jan-12
Art: 714185-00N
The i-STAT
®
Celite
®
Activated Clotting Time test,
Celite
ACT, is a measure of the time required for complete
activation of the coagulation cascade.
1
In traditional ACT tests, coagulation is initiated by mixing a whole blood sample with a particulate activator,
and complete activation is indicated when extensive or localized clots form as activated thrombin converts
fibrinogen to fibrin. These clots are mechanically detected.
The i-STAT
Celite
ACT test is similar to traditional ACT tests except that the endpoint is indicated by the
conversion of a thrombin substrate other than fibrinogen and an electrochemical sensor is used to indicate
the event of this conversion. The substrate used in the electrogenic assay has an amide linkage that mimics
the thrombin-cleaved amide linkage in fibrinogen.
The substrate is H-D-phenylalanyl-pipecolyl-arginine-
p
-amino-
p
-methoxydiphenylamine which has the
structure:
Phenylalanine - Pipecolic acid - Arginine -- NH - C
6
H
4
- NH - C
6
H
4
- OCH
3
Thrombin cleaves the amide bond at the carboxy- terminus of the arginine residue (denoted by the two
dashes) because the bond structurally resembles the thrombin-cleaved amide linkage in fibrinogen. The
product of the thrombin-substrate reaction is the electrochemically inert tripeptide Phenylalanyl - Pipecolyl
- Arginine and the electroactive compound NH
3
+
- C
6
H
4
- NH - C
6
H
4
- OCH
3
. The formation of the
electroactive compound is detected amperometrically, and the time of detection is measured in seconds.
The test reports the Activated Clotting Time (ACT) as a whole blood time (WBT) in seconds.
The i-STAT
Celite
ACT test is calibrated to match the Hemochron Celite FTCA510 using prewarmed tubes.
However, users of the i-STAT
®
1 analyzer may choose to customize their individual i-STAT locations to report
ACT results as calibrated against the Hemochron Celite ACT using non-prewarmed (ambient) temperature
tubes. This customization affects the Patient path only, and will not be applied to the Control or the
Proficiency Testing pathway.
The customization in effect (prewarm or non-prewarm calibration mode) is identified on the analyzer screen
as PREWRM or NONWRM, respectively. Please note that different locations within a given hospital may
utilize different customization profiles. Prior to patient sample testing, ensure the appropriate calibration
mode is employed. For a comprehensive discussion of this customization feature, please see the Technical
Bulletin entitled “ACT Test Result Calibration Options: PREWARMED vs. NON-PREWARMED Result
Calibration Modes for the i-STAT
®
1 Analyzer”.
If results appear inconsistent with the clinical assessment, the patient sample should be re-tested using
another cartridge.
Intended Use
The i-STAT Celite Activated Clotting Time (
Celite
ACT) test is an
in vitro
diagnostic test that uses fresh, whole
blood, and is useful for monitoring patients receiving heparin for treatment of pulmonary embolism or
venous thrombosis, and for monitoring anticoagulation therapy in patients undergoing medical procedures
such as catheterization, cardiac surgery, surgery, organ transplant, and dialysis.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......