9-8
Art: 714371-00E
Rev
. Date: 06-Aug-12
APPLY QC SCHEDULE
TO
The months of the year to which this schedule will apply.
Options are:
All months
Selected Months: Check the box next to the months to which this schedule will
apply.
All months
CARTRIDGE QC
PROFILE
The System Administrator defines a QC cartridge set consisting of:
A QC cartridge type (i.e. the cartridge type to be test with specified liquid QC
fluids during the QC procedure), as well as
Any number of dependent cartridge types (i.e. associated cartridge types that
will be enabled by the handheld if the QC requirements for a given cartridge
set are met on that handheld).
The System Administrator then associates the defined cartridge set with up to
six (6) specific QC fluids.
Disabled
A Cartridge QC Profile can only be edited or created if QC
Frequency has been activated.
See the Technical Bulletin “Liquid Quality Control Schedule
and Lockout Customization on the i-STAT 1 Handheld” for
full details on defining a Cartridge QC Profile.
PREFERENCE WINDOW:
for
r
esults
r
eportIng
o
ptIons
Option
Description
Default
Comments
REFERENCE RANGES
Reference ranges can be defined for each test. The ranges will be
depicted as tic marks on the bar graphs on the result pages. There
are no bar graphs for blood gas, coagulation, and immunoassay
tests.
Ranges are listed in
the Cartridge and Test
Information sheets.
Ranges will be displayed on the Customization screen of the handheld
under the Administration Menu.
Only one range is allowed for each test in a particular handheld.
However, different customization profiles can be set up in specific
handhelds used for specific patient populations.
Care should be taken to enter the same units as selected in the Unit
Set Window.
ACTION RANGES
High and low action ranges can be defined for each test.
Disabled
(-99999.9 to 99999.9)
Care should be taken to enter Action Ranges within the
reportable ranges of the tests.
Care should be taken to enter the same units as selected in
the Unit Set Window.
CUSTOM REPORTABLE
RANGES
High and low custom Reportable Ranges can be defined for each
analyte (except ACT).
Disabled
(-99999.9 to 99999.9)
See the "Reportable Range Customization on the i-STAT 1 Handheld"
Technical Bulletin for full details.
PRINT REFERENCE
RANGES
Reference Ranges can be printed with results. Ranges will print only
if the record to be printed is stored with the active Preference set in
the handheld.
Disabled
The active Preference set in the handheld is listed as ”Custom” on the
Analyzer Status page and the Preference set stored with the record is
displayed on the Chart Page when the record is recalled and is printed
with the results.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......