9-6
Art: 714371-00E
Rev
. Date: 06-Aug-12
PREFERENCE WINDOW:
for
C
artrIdge
QC – e
leCtronIC
QC s
ettIngs
Option
Description
Default
Comments
EXTERNAL SIMULATOR
SCHEDULE
Options are Off (no prompt), an interval of specified hours (1 to 65535 hours), or an
interval of specified patient tests (up to 99999).
The behavior of the analyzer if the schedule is not met can also be specified: Warn or
Lockout (testing disabled until Simulator used).
No prompt
For the quality control of i-STAT analyzers,
i-STAT recommends the use of the Electronic
Simulator.
i-STAT’s recommendation for the frequency
of the Electronic Simulator is once every 24
hours. More frequent use or use according
to number of patient tests may be required
by accreditation and regulatory bodies.
INTERNAL SIMULATOR
SCHEDULE
Time interval when the internal Electronic Simulator test will be run. Options are
Off; an interval of specified hours (1 to 65535 hours); 8/24 (every 8 hours for blood
gases, coagulation, hematocrit and immunoassays, and every 24 hours for other
tests); an interval of specified patient tests (up to 99999).
The behavior of the analyzer if the simulator test fails can also be specified. If the
Schedule Option Lockout is selected, the analyzer will continue to perform the
simulator test and will continue to display “FAIL” on subsequent cartridges until the
test passes. If Lockout is not selected, the simulator test will not be initiated again
until next scheduled time.
Interval 24 hours.
Lockout
PREFERENCE WINDOW:
for
C
artrIdge
QC – l
IQuId
QC s
ettIngs
Option
Description
Default
Comments
CONTROL PASS/FAIL
DETERMINATION
Describes the way in which the System Administrator will determine the acceptability
of liquid QC results.
Options are:
None: Disables the QC Pass/Fail and QC Schedule feature.
Automatic via EVAS: Choosing this option indicates that the handheld will
automatically determine whether the liquid QC run passed or failed, based upon QC
ranges contained on an electronic Value Assignment Sheet (eVAS) file downloaded
into the i-STAT 1 handheld.
Manual: The user will manually compare the liquid QC results to a Value
Assignment Sheet downloaded or printed from the Abbott Point of Care (APOC)
website at www.abbottpointofcare.com/valsheets and indicate on the handheld
whether the QC run passed or failed.
None
See the Technical Bulletins “Liquid Quality
Control Schedule and Lockout Customization
on the i-STAT 1 Handheld” and “Liquid
Quality Control Pass/Fail Customization on
the i-STAT 1 Handheld” for full details.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......