Art: 714260-00N
Rev. Date: 19-Jan-12
2
Code
Number
Cause/Action
Message on Display
Explanation
2
Temperature Out of
Range
/ Check Status
Page
The analyzer is recording a temperature outside its operating
range. Move the analyzer to an area within the operating
temperature of the test being performed and allow the analyzer
to come to the new room temperature. Check the analyzer’s
temperature reading on the Status Page.
3
New Software Installed
/ Use Electronic
Simulator
This message appears on the Portable Clinical Analyzer after
new software has been installed or, in some cases, when a new
customization profile is received.
4, 8
Analyzer Interrupted
/
Use Another Cartridge
The analyzer has detected that the last test cycle was not
completed. This can happen if the batteries were removed
or were making poor contact while a cartridge was still in the
analyzer. Batteries that are too short will not make proper
contact. Check that the batteries are inserted properly and
seated well in the analyzer; check the battery voltage on the
analyzer’s Status Page and replace batteries if low. NOTE:
Patient results displayed before this code are valid.
5, 6, 9
Analyzer Interrupted
/
Ready for Use
The Portable Clinical Analyzer is unable to refresh the display.
This can happen if power is interrupted before the analyzer
powers itself down. Check that batteries are inserted properly
and seated well in the analyzer. Batteries that are too short will
not make proper contact. Check the battery voltage on the
Status Page.
7
Batteries Changed
/
Ready for Use
This is a normal response on the Portable Clinical Analyzer
when the batteries are changed after a code 1 has occurred.
10
Temperature In Range
/
Ready for Use
Temperature is back in range following a code 2 on the Portable
Clinical Analyzer.
11
Date Invalid
/ Check
Clock on Status Page
If the date in the real time clock precedes the release date
programmed into the application software, code 11 is triggered.
Check the date on the real time clock.
The accuracy of the clock is checked at the beginning of
a coagulation test. If the clock is inaccurate, Code 11 is
triggered.
12
Invalid or Expired CLEW
/ See Manual
The CLEW standardization has expired. Download a valid
CLEW.
The date on the real time clock exceeds the expiration date of
the CLEW software. Check the date on the real time clock.
13
Invalid or Expired CLEW
/ See Manual
The CLEW is corrupt or not compatible with the application
software (JAMS), or there is no CLEW in the analyzer.
Download a valid CLEW. If this code occurs after a software
upgrade and the customization application is enabled in the
CDS, change the CLEW version in the Customization Profile to
the latest version and re-transmit the profile to the analyzer.
14
Analyzer Error
/ See
Manual
Customization profile is corrupted. Retransmit the
customization profile. If code 14 reoccurs, contact i-STAT
Technical Services or your local support organization for further
assistance.
15
Barcode Does Not
Match Cartridge Type
The barcode scanned by the user does not match the
immunoassay cartridge type indicated by the identification
chip in the cartridge. The user should run another cartridge,
being careful to scan the barcode from the portion pack of the
specific cartridge type being run on the analyzer.
95
Test Cancelled by
Operator
This message will appear in the stored test records on the
i-STAT 1 Analyzer if the analyzer powers down before
mandatory information was entered.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......