Error in Cartridge
or Fluid Movement
(continued)
Message on Display
Cause
Action
Insufficient Sample
Use Another Cartridge
This is most likely due to
insufficient sample in the
sample well of the cartridge,
but can also be caused by
bubbles in the sample.
Try another cartridge.
Cartridge Not Inserted
Properly
Reinsert Cartridge
The code indicates the
cartridge or external
Electronic Simulator may
not be pushed in all the
way.
Reinsert the cartridge or Electronic
Simulator. If problem is recurrent and/
or the user is certain the cartridge or
Simulator is properly inserted, it may
indicate an instrument problem. Refer
to Support Services.
Test Cancelled by Operator No response to mandatory
prompt before analyzer time
out.
No action required. Training may be
required if a particular operator has a
high rate of cancelled tests.
Electrical or
Mechanical Failures
The following conditions are related to electronic or mechanical failures in
the analyzer.
Message on Display
Cause
Action
Analyzer Error
Use Electronic Simulator
The analyzer usually
recovers from these
errors when the Electronic
Simulator is run. This
error can occur if the
cartridge or Electronic
Simulator was “angled”
when inserted.
Push cartridge or Simulator straight through
the cartridge port. This error can also occur if
the Electronic Simulator is malfunctioning (has
it been dropped?). Try another Simulator. If
the analyzer passes the Electronic Simulator
check, continue to use it. If not, or if the
Quality Check Code is recurrent, the analyzer
may need repair.
Analyzer Error
See Manual
These are mechanical or
electronic failures from
which the analyzer may
not be able to recover.
Use an external Electronic Simulator twice and
use a cartridge with sample or control solution.
If an error condition occurs, refer to Support
Services.
If not, continue to use the analyzer.
Cartridge Type Not Recognized
Use Another Cartridge
This error could be due to
use of a cartridge type that
is not compatible with the
version of software in the
analyzer.
If this is a new cartridge type being used,
update the software. If the cartridge type has
been used before, check to see if the cartridges
have expired. Otherwise, an analyzer problem
is indicated and the analyzer may need repair.
Internal Simulator Failure
This error can occur if
poor contact is made
between the handheld
pins and the contact
pads of the cartridge.
Lockout Enabled:
Immediately rerun
the cartridge in the same handheld. If
the simulator test fails again, rerun the
cartridge is another handheld. Note:
the cartridge should not be run if there
is more than a three minute delay from
the time it was filled. Verify the failed
handheld using an external electronic
simulator.
Lockout Not Enabled:
Immediately
rerun the cartridge in another handheld.
Note: the cartridge should not be run if
there is more than a three minute delay
from the time it was filled. Verify the failed
handheld using an external electronic
simulator.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......