Rev. Date: 01-Aug-11
Art: 714180-00K
PO2 - 3
Method comparison data were collected using CLSI guideline EP9-A
4
. Arterial blood samples were collected
from hospital patients in 3cc blood gas syringes and were analyzed in duplicate on the i-STAT System and
the comparative method within 5 minutes of each other.
Deming regression analysis
5
was performed on the first replicate of each sample. In the method comparison
table, n is the number of specimens in the data set, Sxx and Syy refer to estimates of imprecision based
on the duplicates of the comparative and the i-STAT methods respectively, Sy.x is the standard error of
the estimate, and r is the correlation coefficient.* Method comparisons will vary from site to site due to
differences in sample handling, comparative method calibration and other site specific variables.
* The usual warning relating to the use of regression analysis is summarized here as a reminder: For any analyte, “if the data is collected over
a narrow range, the estimate of the regression parameters are relatively imprecise and may be biased. Therefore, predictions made from these
estimates may be invalid”.
6
The correlation coefficient, r, can be used as a guide to assess the adequacy of the comparative method range in
overcoming this problem. As a guide, the range of data can be considered adequate if r>0.975.
Precision Data (mmHg)
Aqueous Control
Mean
SD
%CV
Level 1
65.1
3.12
4.79
Level 3
146.5
6.00
4.10
Method Comparison (mmHg)
Radiometer
ABL500
Radiometer
ABL700
Bayer 845
n
45
29
30
Sxx
3.70
2.04
3.03
Syy
2.78
2.64
3.28
Slope
1.023
0.962
1.033
Int’t
-2.6
1.2
-2.9
Sy.x
2.52
3.53
3.44
Xmin
----
39
31
Xmax
----
163
185
r
0.996
0.990
0.996
Factors Affecting Results*
Exposure of the sample to air will cause an increase in
P
O
2
when values are below 150 mmHg and a
decrease in
P
O
2
when values are above 150 mmHg (approximate
P
O
2
of room air).
Standing anaerobically at room temperature will decrease pH at a rate of 0.03 per hour, will increase
P
CO
2
by approximately 4 mmHg per hour and will decrease
P
O
2
at a rate of 2–6 mmHg per hour.
6
Do not ice samples before testing –
P
O
2
results may be falsely elevated in cold samples. Do not use a cold
cartridge –
P
O
2
results may be falsely decreased if the cartridge is cold.
*It is possible that other interfering substances may be encountered. These results are representative and your results may differ somewhat due to
test-to-test variation. The degree of interference at concentrations other than those listed might not be predictable.
Factors Affecting Calculated Results
sO
2
values calculated from a measured
P
O
2
and an assumed oxyhemoglobin dissociation curve may differ
significantly from the direct measurement.
3
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......