Activity Versus
Concentration
Ion-selective electrodes measure activity rather than concentration. Activity (a)
is related to concentration (c) through the activity coefficient (
γ
): a =
γ
c.
While ion activities, which reflect free rather than total ion concentrations,
are the physiologically relevant quantity, activity values are converted to
conventional concentration units so that values obtained by direct ISE
measurements can be compared to values obtained from methods that measure
total ion concentrations. The latter includes the indirect methods, which
have activity coefficients close to unity or one, and flame photometric, atomic
absorption and titration methods.
Amperometric
Sensors
In amperometric measurements, a potential is applied to the measuring
electrode while current generated by the resulting oxidation or reduction
reactions in the test system is measured. The current generated is directly
proportional to the concentration of the analyte. An enzyme can be added to a
layer on or near an amperometric sensor to produce electroactive species from
analytes of interest that cannot themselves be oxidized or reduced.
Conductometric
Sensors
In a conductometric measurement, an alternating current is applied between
two electrodes in contact with the test solution and the resulting voltage
difference is measured. The conductivity of the solution is proportional to
the magnitude of the voltage difference. In aqueous solutions, conductivity is
dependent upon the concentration of electrolytes; an increase in the electrolyte
concentration causes an increase in conductivity.
Where S replaces the constant term which defines the slope of the sensor.
The slope is the change in millivolts per tenfold change in the activity of the
analyte. For a positively-charged monovalent ion, the theoretical slope would
be 59.1 mV at 25°C.
DETERMINATION OF TEST RESULTS
Determination
of Analyte
Concentration
Potentiometric and amperometric sensors are used for the determination of
analyte concentration. For both sensors, the concentration of the analyte
can be calculated using:
1) the known value of the analyte concentration in the calibrant
solution,
2) the measured voltage (potentiometric) or current (amperometric)
signal generated by the analyte in the calibrant, and
3) the measured signal generated by the analyte in the test solution.
For potentiometric sensors, the analyte activity in the sample is calculated from
the Nernst equation according to:
E
sample
- E
calibrant
= S log (a
sample
/a
calibrant
).
Complex solutions such as blood deviate slightly from Nernstian behavior due
to interfering ions and matrix effects that result in junction potentials. By
including selectivity coefficients in the Nernst equation (Nikolsky equation),
these effects can be minimized. By characterizing the reference electrode in
different solutions, effects of matrix on the reference junction potential can
also be minimized.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......