15-6
Art: 714377-00K
Rev. Date: 31-Jan-12
Control Target Values
and Ranges
See value assignment sheet posted on the APOC website at
www.abbottpointofcare.com. The value assignment sheet displays target values
and ranges expected when controls and equipment are performing properly.
Always ensure that the lot number and software revision on the value
assignment sheet matches the lot number of the vial in use and the software
revision in the analyzer.
Target values are specific to the i-STAT System. Results may differ if used with
other methods.
i-STAT BNP CALIBRATION VERIFICATION CONTROL (LEVELS 1, 2, AND 3)
Intended Use
i-STAT BNP Calibration Verification Controls are assayed liquid plasma used to
verify the calibration of the BNP test throughout the reportable range.
Warning and
Precautions
Handle the products using the same safety precautions used when handling any
potentially infectious material. The human plasma used in the preparation of
these products has been tested by FDA approved test methods and found negative/
non-reactive for HIV-I, HIV-2, HBsAg, and HCV. However, no known test method
can offer complete assurance that products derived from human blood will not
transmit infectious disease.
Each 1.0 mL vial contains <0.1% sodium azide as a preservative. Dispose of this
product according to local, state, and national regulations. If disposed down a
drain, the sodium azide in this product may react with lead and copper plumbing
to form highly explosive metal azides. Flush drain with large amounts of water
to prevent azide build-up.
Do not use if the calibration verification materials arrive thawed or uncapped.
Bacterial contamination of the control can cause an increase in turbidity. Do not
use the calibration verification material if there is visible evidence of microbial
growth or gross contamination.
Storage and Stability
Store at
≤
-18˚C (-1˚F) in a non-defrosting freezer. After thawing, the opened or
unopened vial is stable for 4 hours when capped and stored at 2-8˚C (35-46˚F).
Do not re-freeze.
Procedure
1. Remove vial from freezer and thaw at room temperature (18-30˚C) for
15 minutes.
2. Gently invert the vial 10 times, then swirl the vial 10 times. Inspect
the sides of the vial to ensure that no particulate matter is clinging to
the sides of the vial. If solids are observed in the control fluid or on
the vial wall, repeat the mixing procedure. If further mixing does not
homogenize the sample, discard the vial and thaw a fresh vial.
3. Open the vial and transfer a drop of solution directly from the vial into
the BNP cartridge and seal the cartridge. If short-term storage (<4 hrs)
is desired, tightly recap the bottle immediately after all sampling is
complete, and store at 2-8˚C.
4. Insert cartridge into an i-STAT 1 Analyzer.
Target Values and
Ranges
See Value Assignment Sheet posted on the APOC website at
www.abbottpointofcare.com. The value assignment sheet displays target values
and ranges expected when controls and equipment are performing properly.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......