Art: 723270-00A
Rev. Date: 20-Apr-12
4
Transfer with Capillary Tube
Plain capillary tubes are recommended to transfer an aqueous control solution from
the ampule to the cartridge. When using a capillary tube (fresh capillary tubes with
sufficient fill capacity are recommended), fill from the bottom of the ampule to avoid
drawing air into the capillary tube. Avoid drawing solution from the surface by placing a
finger over the far end of the tube as it is inserted into the ampule. Once the open end
of the tube rests at the bottom of the ampule, uncover the other end to allow filling by
capillary action.
Transfer with Syringe
Plain syringes (fresh 1cc or 3cc sterile syringe with 16 – 20 gauge needles) are
recommended to transfer aqueous control solutions from the ampule to the cartridge.
When using a syringe, slowly draw approximately 1mL of solution from the bottom of the
ampule.
If air is trapped between the leading edge of the solution and the plunger, do not invert
the syringe to expel it; this will not affect solution near the tip of the syringe.
If air bubbles are continually drawn into the syringe, or if a bubble is trapped near the tip
of the syringe, discard the ampule and syringe and use a fresh ampule and syringe.
Expel one or two drops from the syringe before filling the cartridge.
Target Values
Target values (determined by testing multiple ampules of each level using multiple lots
of cartridges and i-STAT handhelds that have passed the Electronic Simulator test) are
printed on a Value Assignment Sheet posted on the APOC website at
www.abbottpointofcare.com.
Ensure that the lot number printed on the Value Assignment Sheet matches the lot
number on the label of the ampule and that the software version above the target value
table matches the software version in the handheld.
Ranges
The ranges displayed represent the maximum deviation expected when controls and
cartridges are performing properly.
Should results outside these ranges be obtained, refer to the Troubleshooting section
that follows the Procedure for Testing Controls in the System Manual.
Target Values are specific to the i-STAT System. Results obtained when testing these
aqueous controls with other methods may differ due to matrix effects.
Correction of
P
O
2
at Extreme Altitude
The partial pressure of oxygen in a solution will change as it equilibrates to the
surrounding ambient pressure. The rate of change is faster in aqueous solutions than in
whole blood due to the absence of hemoglobin which binds oxygen. This is of practical
significance when testing aqueous solutions on blood gas analyzers as there will be a
detectable shift in the partial pressure of oxygen in the sample as it equilibrates to the
pressure in the flowpath of the analyzer.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......