MAINTAINING THE CERAMIC CONDITIONING CARTRIDGE
The Ceramic Conditioning Cartridge consists of an aluminum base that supports a ceramic “strip.” The
strip is a white strip of Alumina that is held down by a brass retainer and retainer screw. The ceramic
cartridge may be used up to 50 times before the strip is worn and needs to be rotated or up to 200 times
before the strip must be replaced as described below.
Request replacement parts by contacting your i-STAT support provider and refer to the following i-STAT part
numbers:
Description
List/Part Number
Ceramic Cartridge, Packaged (with box, spare strips, etc.)
04J51-01
User Guide, Ceramic Cartridge
014540-01
Ceramic Cartridge (Cartridge only)
014499-01
Ceramic Strip
014542-01
Screw
014543-01
Retainer (brass plate with hole in it)
014544-01
Base (Aluminum support shaped like cartridge)
014545-01
PROCEDURE FOR ROTATING THE STRIP
Note: Rubber gloves should be worn to perform the following procedure.
1. Using a small Phillips head screwdriver, loosen and remove the screw and retainer.
2. Remove the ceramic strip.
Note:
The ceramic strip is brittle and should be handled with care to avoid damaging or
contaminating it.
3. Inspect the ceramic strip for damage. Replace if cracked or chipped. CRACKED STRIPS MUST
BE REPLACED BEFORE USING THE CERAMIC CARTRIDGE IN AN ANALYZER.
4. Inspect the aluminum base. Clean if necessary with isopropyl alcohol and soft, lint-free cloth.
Avoid using paper that might leave fibers on the ceramic cartridge which might be carried into
the analyzer.
5. Rotate the ceramic strip to the next orientation (either spin around or flip over).
Note:
The ceramic cartridge may be used to perform 25 repairs before rotating or replacing
the strip. The strip may be rotated 3 times before replacing it (i.e., the strip has a total of
4 positions; original position of the strip plus 3 rotations). In other words, a single strip
may be used to repair 100 analyzers – 25 analyzers per orientation of the strip. The 4
orientations are:
1. Initial position
2. The strip rotated by “spinning it” 180 degrees, same side up.
3. The strip rotated by turning it over, now back-side up.
4. The strip rotated by “spinning it” again 180 degrees, back-side still up.
6. Place the ceramic strip in the base, making sure it sits flat in the recess and does not overhang
either edge (i.e., the inner edges adjacent to the strip, not the wider edges back on the body of
the cartridge).
7. Replace the retainer plate and retainer screw.
Note:
Over-tightening the screw may damage the threads in the base or crack the ceramic strip.
8. Record the strip rotation in the Use Log.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......