Art: 714364-00M
Rev. Date: 02-Aug-12
2-7
LCD Display and
Backlight
Audible Indicator
The analyzer will beep to indicate:
whenever a key is pressed.
a successful barcode entry.
results are ready.
a Quality Check Message is displayed.
The analyzer can be customized to disable
beeping when a key is pressed or results or
messages are displayed.
Time Out
The analyzer automatically turns off after a
certain period of inactivity.
Results displayed: Results are displayed for 2 minutes before the analyzer
turns off provided that a mandatory Comment Code prompt is not displayed.
This Inactivity Time Out default time can be increased using Customization.
If a mandatory Comment Code prompt is displayed, the analyzer will turn
off after 15 minutes or after the Inactivity Time Out, whichever is greater. In
the case of a missed required Comment Code, results will be stored and “_
_ _” will be entered as the Comment Code.
Prompting for mandatory data when results are ready for display:
The analyzer will turn off after 15 minutes or after the Inactivity Time Out,
whichever is greater, if there is no response to a mandatory data prompt.
A mandatory data prompt is a prompt for information that must be entered
before pending results are displayed.
In the case of a missed mandatory data prompt, results will not be stored
and the test record will state “Test Cancelled by Operator.”
Waiting for insertion of cartridge: After the prompt “Insert Cartridge”
is displayed, the analyzer will wait 15 minutes for the operator to insert a
cartridge unless the analyzer is in the Proficiency path, in which case the
analyzer will wait 5 minutes. If a cartridge is not inserted, the analyzer will
turn off. This timeout cannot be customized.
Other: The analyzer will turn off after 2 minutes of inactivity (no keys pressed)
in all other circumstances.
Test results, operator prompts and other messages are displayed on the analyzer’s
LCD Screen. The backlight for the display is turned on and off by pressing the 0
key for one second. The backlight will automatically turn off after ninety seconds
and when the analyzer powers down or is turned off. The backlight cannot be
turned on while data entry screens are displayed.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......