PROCEDURE MANUAL FOR THE i-STAT SYSTEM
20
REV. DATE: 16-Oct-12
ART: 714446-00L
Integrity Testing
From each lot of blood gas/chemistry cartridges received, use a representational number of cartridges to analyze i-STAT
or TriControls Level 1 and 3 Controls. For CHEM8+ cartridges, analyze i-STAT CHEM8+ or TriControls Control Levels
1 and 3. If not using TriControls and hematocrit is included on the cartridge type being tested, also analyze RNA Medical
Hematocrit Control Levels 1 and 3. For ACT cartridges, analyze i-STAT Level 1 and Level 2 ACT Controls. For
PT/INR cartridges, analyze i-STAT Level 1 and Level 2 PT Controls. For cTnI cartridges, analyze i-STAT Level 1 and
Level 3 Cardiac Marker Controls. For CK-MB cartridges, analyze CLINIQA Cardiac Marker Level 1 and 3 controls. For
BNP cartridges, analyze i-STAT, BNP Control Levels 1 and 3. Use any verified analyzer for control testing. Transmit
the results to the Central Data Station. Use the expected values published in the Value Assignment Sheet to verify the
integrity of the cartridges.
Procedure for testing cartridges with i-STAT Level 1 and Level 3 Controls:
1.
Prior to testing cartridges that measure
P
O2, ampules should stand at room temperature a minimum of 4 hours before
use. When testing other cartridges (G, Crea, E3+, EC4+, 6+, or EC8+), ampules may be used once the fluid has
reached room temperature, approximately 30 minutes for individual ampules. For best results, ampules, cartridges, and
handhelds should be at the same temperature. When using cartridges that contain sensors for measuring ionized
calcium, pH,
P
CO
2
, or
P
O
2
(G3+, EG6+, EG7+, CG4+, CG8+, or EC8+), a separate ampule must be used for each
cartridge being tested; if these sensors are not present (i.e., the 6+ cartridge), the contents of one ampule may be used
to fill more than one cartridge as long as the cartridges are filled and inserted into a handheld within 10 minutes of
opening the ampule.
2.
Immediately before use, shake the ampule vigorously for 5 to 10 seconds to equilibrate the liquid and gas phases. To
shake, hold the ampule at the tip and bottom with forefinger and thumb to minimize increasing the temperature of the
solution. If necessary, tap the tip of the ampule to send solution back into the bottom section of the ampule. Protect
fingers with gauze, tissue, or glove, or use an ampule breaker to snap off the tip of the ampule at the neck.
3.
Immediately transfer the solution from the ampule into a plain capillary tube or plain syringe, and then immediately
transfer the solution into a cartridge. Immediately seal the cartridge and insert it into a handheld. It is important not to
expose the solution to room air since this will alter the results.
When using a capillary tube, fill from the bottom of the ampule. Avoid drawing solution from the surface by
covering the far end of the tube as it is inserted into the ampule. Once the open end of the tube rests at the bottom
of the ampule, uncover the other end to allow filling by capillary action.
When using a syringe (1cc or 3cc syringes with 16 to 20 gauge needles are recommended), slowly draw
approximately 1 mL of solution from the bottom of the ampule. If air is trapped between the leading edge of the
solution and the plunger, do not invert the syringe to expel it; this will not affect solution near the tip of the
syringe. If air bubbles are continually drawn into the syringe, or if a bubble is trapped near the tip of the syringe,
discard the ampule and syringe and use a fresh ampule and syringe. Expel one or two drops from the syringe
before filling the cartridge.
Do not use solution left in the syringe, ampule, or capillary tube for additional testing of the cartridges that
contain sensors for ionized calcium, pH,
P
CO
2
, or
P
O
2
. However, cartridges without these sensors may be tested
with remaining fluids if within 10 minutes of opening the ampule.
4.
Compare results to the value assignment sheet ranges. Check that the lot number on the control ampule matches the lot
number on the package insert and that the software version listed on the insert matches the software installed in the
handheld. If all results are within expected ranges, use the cartridges as needed. Transmit the results to the Central Data
Station.
Procedure for testing cartridges with TriControls Level 1 and Level 3 Controls:
1
Prior to testing cartridges that measure
P
O2, ampules should stand at room temperature a minimum of 4 hours before
use. When testing other cartridges (G, Crea, E3+, EC4+, 6+, EC8+, or CHEM8+), ampules may be used once the fluid
has reached room temperature, approximately 30 minutes for individual ampules. For best results, ampules, cartridges,
and handhelds should be at the same temperature. When using cartridges that contain sensors for measuring ionized
calcium, pH,
P
CO
2
, or
P
O
2
(G3+, EG6+, EG7+, CG4+, CG8+, EC8+, or CHEM8+), a separate ampule must be used
for each cartridge being tested; if these sensors are not present (i.e., the 6+ cartridge), the contents of one ampule may
be used to fill more than one cartridge as long as the cartridges are filled and inserted into a handheld within 10
minutes of opening the ampule.
2
Immediately before use, shake the ampule vigorously for 5 to 10 seconds to equilibrate the liquid and gas phases. To
shake, hold the ampule at the tip and bottom with forefinger and thumb to minimize increasing the temperature of the
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......