Art: 714261-00E
Rev. Date: 10-Jan-11
4
White Blood Cells
Interference from white blood cells depends upon the size of the cell. As an example, the hematocrit reading
will be increased by 1 %PCV when the white cell count is 50,000 per microliter, assuming the average white
blood cell occupies twice the volume of an average red blood cell.
Automated Hematology Methods
Osmolality:
16-27
Abnormal sample osmolality can create discrepant readings as the red cells shrink or swell to achieve
osmotic balance with the isotonic (normal osmolality) diluent. Differences up to 10 %PCV have been
reported for glucose concentrations in the range of 1000 mg/dL to 2000 mg/dL. Differences up to 5 %PCV
have been reported for samples with low or high sodium concentrations.
Cold Agglutinins:
28-32
Under the right conditions of diluent temperature, red blood cells from certain samples will clump together.
The large clumps will not be recognized as red blood cells by certain automated hematology analyzers, and
thus the overall red blood cell count and the calculated hematocrit, will be falsely low. In extreme cases, the
hematocrit can read low by up to about 8 %PCV.
Mean Cell Hemoglobin Concentration:
11-15
Red blood cells can deform under the pressure experienced when being sent through the measuring orifice.
This is particularly true for electrical impedance methods. Cells with lower hemoglobin concentrations will
deform to a greater degree. This can create errors as large as 3 %PCV.
Early authors attributed the discrepancy between automated hematology determinations and reference
microhematocrit determinations to variabilities in trapped plasma on the microhematocrit determination.
5-9,10a,10b
Recent authors have recognized the substance of the discrepancy to be an interference on the automated
method.
Co-oximetry Methods
The accuracy and precision of measured total hemoglobin from co-oximeters depends on the instrument’s
efficiency in hemolyzing the blood and the number of wavelengths used to determine hemoglobin. The
calculation of hematocrit from measured hemoglobin assumes a normal MCHC. Deviation from a normal
adult MCHC, e.g. in children, may give an inaccurate calculated hematocrit.
41
Other limitations of measuring
total hemoglobin by spectrophotometric methods include the effect of sulfhemoglobin and fetal hemoglobin
and hyperlipidemic plasma that may falsely elevate the hemoglobin.
42,43
An extremely high number of WBCs
may also elevate the hemoglobin value resulting in an inaccurately calculated hematocrit.
44
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......