DETERMINATION OF COAGULATION ENDPOINTS
ACT and PT/INR
In coagulation tests, the result that is reported is the time required for the
process of coagulation to occur. To determine this time, there must be a
detectable change in a sample parameter correlated to progression of the
process. In traditional coagulation tests, endpoint detection typically relies
on monitoring increases in either blood viscosity or plasma turbidity that
occurs as thrombin converts fibrinogen to clotable fibrin. In an electrogenic
test an electroactive marker that can be detected at either an amperometric or
potentiometric sensor is used to indicate the endpoint. The marker is generated
when a substrate that has been added to the test sample is acted upon by
thrombin. As the coagulation reaction proceeds, the marker concentration
increases, increasing the signal at the sensing electrode. The time required
for generation of the marker correlates to the time required for conversion
of fibrinogen. The coagulation endpoint can, therefore, be determined by
monitoring the marker concentration. Unlike traditional coagulation tests,
electrogenic tests will not be prolonged in samples with abnormally low (less
than 100 mg/dL) fibrinogen levels.
QUALITY CONTROL AND THE I-STAT SYSTEM
Overview
Quality control, as a component of an overall quality assurance program,
consists of tests and procedures for monitoring and evaluating the analytical
performance of a measurement system to assure the reliability of patient test
results.
As new technologies evolve, quality control regimens must match the
requirements of the particular analytical system. i-STAT Corporation
recognizes the importance of effective quality control for its analytical
medical devices, and has developed a program that is tailored to the unique
characteristics of the i-STAT System.
The i-STAT System performs blood analysis when a unit-use cartridge filled
with a patient’s sample is inserted into a handheld analyzer or Blood Analysis
Module (as part of the Agilent Viridia Patient Monitoring System).
The measurement methodologies are electrochemical, using microfabricated
sensors housed in each cartridge to measure analyte concentrations directly in
a single whole blood sample (i.e., neither dilution nor reagent mixing steps are
required).
Two characteristics of the i-STAT System, which distinguish it from traditional
laboratory equipment, have significant impact upon the design of the quality
control regimen: its intended user and the unit-use cartridge technology.
As the system is intended to be used by individuals not trained in laboratory
science, the onus is upon the system’s design to ensure that the quality of
results is not dependent upon either user technique, skilled maintenance and
calibration procedures, or the accompanying quality control regimens which
ensure these procedures have been properly performed.
The use of unit-use cartridges frees the i-STAT System from these skilled
maintenance and calibration procedures. It also allows for the design of a
quality control system which automatically monitors those aspects of the
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......