3
Art: 714260-00N
Rev. Date: 19-Jan-12
The following codes are associated with the cartridge or fluid movement within a cartridge. These
conditions can be operator or sample related. In most cases, a new cartridge must be used. If a
condition persists, especially if isolated to one analyzer, there may an analyzer problem.
Code
Number
Cause/Action
Message on Display
Explanation
19
No Clot Detected
/ See
Manual
During the PT/INR cycle, no clot was detected. Run another
cartridge. If code 19 reappears, run the sample on an alternate
methodology.
22, 25
Cartridge Error
/ Use
Another Cartridge
These codes occur only for coagulation cartridges if the mixing
of the sample and reagent is compromised. This can be
caused by an insufficient or clotted sample, or by air bubbles in
the sample.
24
Cartridge Error
/ Use
Another Cartridge
The electrical resistance of the calibrant fluid (Rcal) used to
verify the electrolyte concentration is out of specification. This
could occur if the calibrant pack was ruptured well before
the test allowing evaporation to result in a higher electrolyte
concentration.
Besides the electrolyte concentration, the Rcal is also
affected by the temperature and the height and width of the
fluid segment over the conductometric sensor. The analyzer
accounts for the temperature, but the height and width of
the fluid segment can vary from cartridge lot to cartridge lot.
The analyzer has been programmed to compensate for these
lot-to-lot differences by maintaining a running average of the
Rcal values measured from the most recent cartridge runs.
Occasionally, the difference between the Rcal values for two
cartridge lots is large enough to cause the introduction of a new
lot to trigger code 24 on the first few cartridge runs. The Code
24 errors should disappear as the running average adjusts.
However, if code 24 persists after more than 3 cartridge runs on
each analyzer, contact i-STAT Technical Services or your local
support organization.
26
Cartridge Error
/ Use
Another Cartridge
This code occurs if there was a coagulation specific quality
check failure: premature substrate activation, abnormally low
levels of substrate, or invalid fluid motion.
20, 27-29,
32, 33, 40,
41, 45, 87
Cartridge Error
/ Use
Another Cartridge
These codes identify problems with the cartridge such as:
calibrant fluid arriving too soon, too late, or not at all, or noise
in the calibrant fluid signals. Codes 20, 27, 41, and 87 can
be caused by poor contact that can sometimes be corrected
by conditioning the pins in the analyzer using the ceramic
conditioning cartridge. The specific conditioning procedure is
described at the end of this bulletin.
The rate of quality check code 45 can be elevated when
cartridges are run without allowing sufficient time for the
cartridges to equilibrate to room temperature. To minimize the
number of quality check codes, review i-STAT cartridge storage
conditions and allow sufficient time for refrigerated cartridges
to equilibrate to room temperature.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......