BNP - 8
Art: 716969-00J
Rev. Date: 23-May-12
The i-STAT BNP Calibrators are traceable to an internal reference standard that has been prepared
gravimetrically with synthetic BNP. The internal reference standard underwent a one-time value assignment
to align with the ARCHITECT BNP assay with a decision threshold of 100 pg/mL.
An age-matched analysis of the heart failure and non-heart failure populations was performed based on the
data published by the American Heart Association in the 2000 Heart and Stroke Statistical Update
24
and
according to the age structure of the United States population.
25
The age distributions in the intended use
population are approximately as follows: individuals less than 45 years old comprise 9%, individuals 45-54
years old comprise 11%, individuals 55-64 years old comprise 22%, individuals 65-74 years old comprise
26%, and individuals 75 years and older comprise 32%. The resulting combined AUC is 0.87 (0.85 to 0.90,
95% CI).
The clinical sensitivity and specificity using a decision threshold of 100 pg/mL is presented in the following
table.*
Males (Age Group)
All
<45
Years
45-54
Years
55-64
Years
65-74
Years
75+
Years
Sensitivity
71.0%
(328/462)
47.1%
(8/17)
57.1%
(24/42)
57.3%
(51/89)
70.6%
(115/163)
86.1%
(130/151)
95% Confidence Interval
66.6 to
75.1%
23.0 to
72.2%
41.0 to
72.3%
46.4 to
67.7%
62.9 to
77.4%
79.5 to
91.2%
Specificity
94.8%
(403/425)
97.2%
(104/107)
100.0%
(71/71)
97.9%
(92/94)
88.7%
(102/115)
89.5%
(34/38)
95% Confidence Interval
92.3 to
96.7%
92.0 to
99.4%
94.9 to
100.0%
92.5 to
99.7%
81.5 to
93.8%
75.2 to
97.1%
Females (Age Group)
All
<45
Years
45-54
Years
55-64
Years
65-74
Years
75+
Years
Sensitivity
80.5%
(186/231)
44.4%
(4/9)
73.3%
(11/15)
50.0%
(13/26)
80.6%
(58/72)
91.7%
(100/109)
95% Confidence Interval
74.8 to
85.4%
13.7 to
78.8%
44.9 to
92.2%
29.9 to
70.1%
69.5 to
88.9%
84.9 to
96.2%
Specificity
88.4%
(411/465)
95.9%
(94/98)
90.7%
(68/75)
89.6%
(69.77)
85.7%
(114/133)
80.5%
(66/82)
95% Confidence Interval
85.1 to
91.2%
89.9 to
98.9%
81.7 to
96.2%
80.6 to
95.4%
78.6 to
91.2%
70.3 to
88.4%
* Representative data, results in individual laboratories may vary from these data.
Performance Characteristics
Precision data were collected as follows: Duplicates of each control were tested daily for a period of
20 days for each of 3 lots of cartridges, resulting in a total of 434 replicates. The average statistics are
presented below.
Whole blood imprecision data were collected as follows: whole blood samples from 5 healthy donors
were spiked to low, intermediate and high BNP concentrations affording 15 samples, each of which was
measured in 10 i-STAT BNP cartridges from a single cartridge lot; three lots of cartridges were employed.
The mean within-sample BNP concentration ranged from 84 – 3925 pg/mL and the within-sample
imprecision (%CV) ranged from 3.4 to 9.4%; the average BNP concentration and imprecision were 1464
pg/mL and 6.5% respectively.
Summary of Contents for i-STAT 1
Page 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Page 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Page 44: ......
Page 90: ......
Page 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Page 162: ......
Page 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Page 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Page 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Page 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Page 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Page 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Page 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Page 192: ......
Page 194: ......
Page 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Page 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Page 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Page 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Page 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Page 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Page 226: ......
Page 236: ......
Page 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Page 254: ......
Page 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Page 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Page 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Page 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Page 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Page 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Page 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Page 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Page 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Page 536: ......
Page 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Page 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Page 572: ...Art 715617 00D Rev Date 06 11 08 ...
Page 578: ...Art 714547 00F Rev Date 06 11 08 ...
Page 590: ...Art 720735 00B Rev Date 06 11 08 ...
Page 592: ...Art 721106 00B Rev Date 06 11 08 ...
Page 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Page 610: ...Art 720654 00B Rev Date 06 11 08 ...
Page 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Page 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Page 618: ...Art 721296 00A Rev Date 07 16 08 ...
Page 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Page 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Page 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Page 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Page 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Page 658: ......